Finished Goods Box of 50 Inserts and Scanning Labels for Medivac Guardian 2L Catalog Number: FG 12010...
Stryker Corporation began a recall of Finished Goods Box of 50 Inserts and Scanning Labels for Medivac Guardian 2L Catalog Number: FG 12010. Intended adjunct in the estimation of blood loss on . FDA classified it as Class II and gives the reason as "Triton Canister Software, Insert & Scanning Label may cause the Triton Canister System to output inaccurate blood loss estimates can lead to confusion and could potentially result in either delayed recognition or...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1184-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Stryker Corp.
- Product
- Finished Goods Box of 50 Inserts and Scanning Labels for Medivac Guardian 2L Catalog Number: FG 12010. Intended adjunct in the estimation of blood loss.
- Product code
- PBZ Image Processing Device For Estimation Of External Blood Loss
- Reason for recall
- Triton Canister Software, Insert & Scanning Label may cause the Triton Canister System to output inaccurate blood loss estimates can lead to confusion and could potentially result in either delayed recognition or prolongation of postpartum hemorrhage (PPH), or require more aggressive treatment for PPH.
- Root cause tag
- labeling and UDI
- FDA root cause
- Under Investigation by firm
- Action
- Stryker issued Urgent Medical Device Correction letter to Risk Manager, OR Director, Materials Manager dated Jan 25, 2023. Letter states reason for recall, health risk and action to take: Steps: 1. Immediately check your internal inventory to locate any affected Finished Goods Inserts and Scanning labels (FG 12010 and FG 12022). Quarantine and discontinue use of all affected products. 2. A Stryker representative will contact your facility to assist updating the Triton Canister software. 3. Return the enclosed Business Reply Form (BRF) even if you don t have any affected product on hand, via email to [email protected]. 4. If the BRF indicates that recalled product is currently on hand, a shipping label will be provided to return recalled product. Upon receipt of the recalled product, Stryker will arrange replacements. 5. Maintain awareness of this communication internally until all required actions have been completed within your facility. Contact: Sylvia Page | Stryker Instruments Recall Coordinator 269-501-8375 |[email protected]
- Quantity in commerce
- 101 units
- Distribution
- US Nationwide distribution.
- Product codes and lots
- UDI-DI: 10859506006125 Lot Numbers: M2110004, M2201002, M2202001, M2203002, M2206001
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1184-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1185-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1186-2023 | Class II | Ongoing | Voluntary: Firm initiated |