Z-1147-2023Class I

Cardiosave Hybrid Model Nos. 0998-00-0800-31 0998-00-0800-32 0998-00-0800-33 0998-00-0800-34 0998-00-0800-35...

Datascope Corp. / initiated 2023-02-07 / Open, Classified / root cause: software

Datascope Corp. began a recall of Cardiosave Hybrid Model Nos. 0998-00-0800-31 0998-00-0800-32 0998-00-0800-33 0998-00-0800-34 0998-00-0800-35 0998-00-0800-45 0998-00-0800-52 0998-00-0800-53... on . FDA classified it as Class I and gives the reason as "An unexpected shutdown of the IABP may occur due to loss of communication between the Executive Processor PCBA and the Video Generator PCBA. This may result in an interruption to therapy which may threaten the...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-1147-2023
Classification
Class I
Status
Open, Classified
Initiated
Posted
Recalling firm
Datascope Corp.
Product
Cardiosave Hybrid Model Nos. 0998-00-0800-31 0998-00-0800-32 0998-00-0800-33 0998-00-0800-34 0998-00-0800-35 0998-00-0800-45 0998-00-0800-52 0998-00-0800-53 0998-00-0800-55 0998-00-0800-65 0998-UC-0800-31* 0998-UC-0800-33* 0998-UC-0800-52* 0998-UC-0800-53* 0998-00-0800-55* * -- not distributed in US
Product code
DSP System, Balloon, Intra-Aortic And Control
Reason for recall
An unexpected shutdown of the IABP may occur due to loss of communication between the Executive Processor PCBA and the Video Generator PCBA. This may result in an interruption to therapy which may threaten the hemodynamic stability of the supported patient as the user is left unaware to the status of the Cardiosave IABP.
Root cause tag
software
FDA root cause
Software design
Action
On February 7, 2023, the firm notified customers of 4 issues affecting the IABP via an urgent Medical Device Correction letter, which included the present issue. Users were instructed to do the following: 1. If you experience an unexpected shutdown of Cardiosave IABP during therapy, or a frozen or black screen, use another IABP to continue therapy. Until an alternative IABP is located you may attempt to restart the IABP. If the IABP remains non-operational, immediately remove from the patient care environment for further product evaluation. 2. If your device remains inoperable, please contact your service representative to identify the cause and take the necessary actions required. The firm is developing a software correction to address the issue. Once available, a Datascope/Getinge service representative will contact you to schedule the installation of the updated software. This work will be done at no cost to your facility. If you have any questions, please contact your Datascope/Getinge representative or call the Datascope/Getinge Technical Support at 1-888-943-8872, options 4, 2, 1, Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone).
Quantity in commerce
4793 (US); 4407 (OUS)
Distribution
Worldwide Distribution: US (nationwide) including states of: AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, NV, OH, OK, OR, PA, SC, SD, TN, TX, VA, WA, WI, and WV; and OUS (Foreign) countries of: ALBANIA, ALGERIA, ANGOLA, ARGENTINA, AUSTRALIA, AZERBAIJAN, BAHRAIN, BANGLADESH, BARBADOS, BELRUS, BELGIUM, BOSNIA AND HERZEGOVINA, BRAZIL, BRUNEI, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, C¿te d'Ivoire, CROATIA, CZECH REPUBLIC, ECUADOR, EGYPT, EL SALVADOR, ESTONIA, FINLAND, FRANCE,GERMANY,GHANA,GUATEMALA,HONG KONG,HUNGARY,ICELAND,INDIA,INDONESIA IRAN, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KENYA, KOSOVO, KUWAIT, LATVIA, LEBANON, LIBYA, LITHUANIA, LUXEMBOURG, MACEDONIA, MALAYSIA, MEXICO, MOROCCO, MYANMAR, NETHERLANDS, NEW ZEALAND, NIGERIA, NORWAY, OMAN, PAKISTAN, PALESTINE, PERU, PHILIPPINES, POLAND, PORTUGAL, QATAR, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SRI LANKA, SWEDEN, SWITZERLAND, SYRIA, TAIWAN, TANZANIA, THAILAND, TRINIDAD AND TOBAGO TUNISIA, TURKEY, TURKMENISTAN, UKRAINE, UNITED ARAB EMIRATES, UNITED KINGDOM UZBEKISTAN, VIETNAM, YEMEN AND ZAMBIA.
Product codes and lots
Model UDI 0998-00-0800-31 10607567109053 0998-00-0800-32 10607567111117 0998-00-0800-33 10607567109008 0998-00-0800-34 10607567111940 0998-00-0800-35 10607567109107 0998-00-0800-45 10607567108421 0998-00-0800-52 10607567108438 0998-00-0800-53 10607567108391 0998-00-0800-55 10607567108414 0998-00-0800-65 10607567113432 0998-UC-0800-31* N/A 0998-UC-0800-33* N/A 0998-UC-0800-52* N/A 0998-UC-0800-53* N/A 0998-00-0800-55* N/A All serial numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1147-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1147-2023Class IOngoingVoluntary: Firm initiated
Z-1148-2023Class IOngoingVoluntary: Firm initiated