Z-1225-2023Class II

CardioMEMS HF System PA Sensor and Delivery System, Model Number CM3100, The CardioMEMS HF System provides...

Abbott Medical / initiated 2023-02-07 / Open, Classified / root cause: design

St. Jude Medical began a recall of CardioMEMS HF System PA Sensor and Delivery System, Model Number CM3100, The CardioMEMS HF System provides pulmonary artery (PA) hemodynamic data used for the... on . FDA classified it as Class II and gives the reason as "Select CardioMEMS PA Sensors (Model CM2000) operate outside of the intended frequency range at higher elevations when readings are taken over approximately 2,000 ft. (610 meters) above sea level; operation outside of...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-1225-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Abbott Medical
Product
CardioMEMS HF System PA Sensor and Delivery System, Model Number CM3100, The CardioMEMS HF System provides pulmonary artery (PA) hemodynamic data used for the monitoring and management of heart failure (HF) patients. The system measures changes in PA pressure which physicians use to initiate or modify heart failure treatment.
Product code
MOM System, Hemodynamic, Implantable
Reason for recall
Select CardioMEMS PA Sensors (Model CM2000) operate outside of the intended frequency range at higher elevations when readings are taken over approximately 2,000 ft. (610 meters) above sea level; operation outside of the intended radiofrequency has the potential to result in inaccurate readings or sensor signal acquisition difficulties.
Root cause tag
design
FDA root cause
Device Design
Action
Abbott notified consignees on about 02/07/2023 via hand-delivery of the letter within the US. The letter instructed consignees to do the following: 1. Prior to the implantation procedure, sensor data should be entered, and the clinician should proceed to the next screen to complete the Error #8 screening, prior to venipuncture. 2. If an Error #8 message occurs, the sensor should not be implanted. 3. Work with your Abbott Sales Representative to exchange the affected sensor. 4. Review Appendix B for labeling and supplemental information related to implantation and monitoring of CardioMEMS PA Sensors.
Quantity in commerce
317 units
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Canada, Colombia, Czech Republic, Denmark, Finland, France, Germany, Ireland, Italy, Kuwait, Netherlands, Portugal, Saudi Arabia, Spain, Sweden, Switzerland, UK, United Arab Emirates.
Product codes and lots
UDI/DI 05415067034724: M220200004, M220200005, M220200006, M220200008, M220200011, M220200014, M220200016, M220200017, M220200018, M220200019, M220200021, M220200022, M220200026, M220200028, M220200029, M220200031, M220200032, M220200034, M220200035, M220200036, M220200037, M220200039, M220200040, M220200041, M220200042, M220300002, M220300003, M220300004, M220300005, M220300007, M220300008, M220300009, M220300010, M220300011, M220300012, M220300013, M220300014, M220300016, M220300017, M220300018, M220300019, M220300021, M220300022, M220300023, M220300024, M220300025, M220300026, M220300027, M220300028, M220300029, M220300030, M220300031, M220300032, M220300035, M220300036, M220300040, M220300041, M220300042, M220300043, M220300044, M220300045, M220300046, M220300047, M220300048, M220300049, M220300050, M220300051, M220300052, M220300053, M220300054, M220300055, M220300057, M220300058, M220300059, M220300060, M220300061, M220300062, M220300063, M220300064, M220300065, M220300066, M220300067, M220300069, M220300070, M220300071, M220300073, M220300074, M220300075, M220300077, M220300078, M220300080, M220300081, M220300082, M220300083, M220300084, M220300085, M220300087, M220300088, M220300089, M220300091, M220300092, M220300093, M220300096, M220300097, M220300101, M220300102, M220300103, M220300104, M220300105, M220300107, M220300108, M220300109, M220300111, M220300112, M220300114, M220300115, M220300116, M220300118, M220300119, M220300120, M220300121, M220300122, M220300123, M220300125, M220300126, M220300127, M220300129, M220300131, M220300136, M220300138, M220300139, M220300140, M220300141, M220300142, M220300143, M220300144, M220300145, M220300146, M220300147, M220300150, M220300151, M220300152, M220300153, M220300154, M220300155, M220300156, M220300157, M220300158, M220300160, M220300161, M220300162, M220300163, M220300164, M220300166, M220300167, M220300168, M220300169, M220300170, M220300171, M220300172, M220300173, M220300174, M220300175, M220300176, M220300177, M220300178, M220300179, M220300180, M220300181, M220300182, M220300183, M220300184, M220300186, M220300187, M220300188, M220300189, M220300190, M220300191, M220300192, M220300193, M220300194, M220300195, M220300196, M220700025, M220700026, M220700027, M220700028, M220700031, M220700033, M220700034, M220700035, M220700036, M220700037, M220700041, M220700042, M220700043, M220700045, M220700051, M220700052, M220700054, M220700059, M220700061, M220700097, M220700098, M220700102, M220700104, M220700106, M220700107, M220700109, M220700110, M220700111, M220700112, M220700113, M220700114, M220700115, M220700116, M220700117, M220700118, M220700119, M220700120, M220700121, M220700122, M220700123, M220700124, M220700125, M220700126, M220700128, M220700129, M220700130, M220700131, M220700132, M220700133, M220700134, M220700135, M220700136, M220700137, M220700138, M220700139, M220700142, M220700143, M220700144, M220700145, M220700146, M220700147, M220900017, M220900018, M220900021, M220900023, M220900024, M220900025, M220900026, M220900030, M220900031, M220900032, M220900033, M220900034, M220900035, M220900037, M220900038, M220900040, M220900042, M220900044, M220900045, M220900046, M220900047, M220900048, M220900049, M220900050, M220900052, M220900063, M220900065, M220900066, M220900067, M220900068, M220900069, M220900070, M220900071, M220900072, M220900073, M220900075, M220900076, M220900078, M220900079, M220900080, M220900081, M220900086, M220900087, M220900088, M220900089, M220900090, M220900091, M220900093, M220900094, M220900095, M220900096, M220900097, M220900101, M220900104, M220900106, M220900109, M221000023, M221000030, M221000032, M221000034, M221000035, M221000038, M221000039, M221000041, M221000042, M221000044, M221000047, M221000048, M221000049, M221000050, M221000052, M221100002, M221000046

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1225-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1223-2023Class IIOngoingVoluntary: Firm initiated
Z-1224-2023Class IIOngoingVoluntary: Firm initiated
Z-1225-2023Class IIOngoingVoluntary: Firm initiated