Z-1213-2023Class II

Baxter MiniCap Disconnect Cap with Povidone-Iodine Solution, 5C4466P; Accessory to disposable administration...

Baxter Healthcare Corporation / initiated 2023-02-01 / Open, Classified / root cause: manufacturing process

Baxter Healthcare Corporation began a recall of Baxter MiniCap Disconnect Cap with Povidone-Iodine Solution, 5C4466P; Accessory to disposable administration set for peritoneal dialysis on . FDA classified it as Class II and gives the reason as "MiniCap Disconnect Cap with Povidone-Iodine are packaged in foil pouches, which may have been incorrectly sealed, i.e., the pouches may have open or weak seals. This could lead to exposure to air, resulting in...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1213-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Baxter Healthcare Corporation
Product
Baxter MiniCap Disconnect Cap with Povidone-Iodine Solution, 5C4466P; Accessory to disposable administration set for peritoneal dialysis
Product code
KDJ Set, Administration, For Peritoneal Dialysis, Disposable
Reason for recall
MiniCap Disconnect Cap with Povidone-Iodine are packaged in foil pouches, which may have been incorrectly sealed, i.e., the pouches may have open or weak seals. This could lead to exposure to air, resulting in insufficient iodine/dry sponge inside the MiniCap, which could lead to the potential for inadequate disinfectant.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Baxter Healthcare initiated the recall by issuing an Urgent Medical Device Recall notice to its consignees by letter on 02/01/2023. The notice requested the following from customers: 1. Check your stock and set aside all product with the affected lot numbers. 2. Contact Baxter HomeCare Services to arrange for return of impacted product and for ordering replacement product. 3. If all of your product is impacted by this recall and you are unable to get unaffected lot numbers, resulting in a significant delay in therapy, please follow the below steps: a. Contact your clinic for clinical guidance and check each pouch and do not use the MiniCap contained inside of any pouches that appear to have damaged seals. For general questions regarding this communication, contact Baxter Corporate Product Surveillance at 800-437-5176, between the hours of 8:00 am and 5:00 pm Central Time, Monday through Friday." Recall was expanded and Baxter Healthcare sent an updated notification to customers and patients on 05/09/2023. The letter instructed consignees and patients to immediately cease use of the affected product, arrange for the return of affected product and assess each patient's individual next steps based on clinical evaluation. Customers were also instructed to complete the customer response form via the portal (for products directly received from Baxter) or through the distributor. If the products were further distributed, please notify customers.
Quantity in commerce
33,866,400 units
Distribution
US Nationwide distribution and Canada.
Product codes and lots
UDI/DI 00085412007694, Lot numbers: GD912051, exp 31-Mar-2024; GD912068, exp 30-Apr-2024; GD912099, exp 30-Apr-2024; GD912112, exp 30-Apr-2024; GD912143, exp 30-Apr-2024; GD912204, exp 31-May-2024. Expansion on 05/10/2023: Product Code 5C4466P with all lots numbers manufactured prior to 09/30/2022.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1213-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1213-2023Class IIOngoingVoluntary: Firm initiated