BIOPHEN Protein C 5 kit is for in vitro quantitative determination of Protein C activity on human citrated...
Aniara Diagnostica LLC began a recall of BIOPHEN Protein C 5 kit is for in vitro quantitative determination of Protein C activity on human citrated plasma. Protein C is a glycoprotein, vitamin K... on . FDA classified it as Class II and gives the reason as "New warnings and precautionary statements on product labelling: Reagent 2 (R2) of BIOPHEN Protein C 5 contains the chemical substance - Cesium Chloride which changed classification and has become hazardous according to...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1227-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Aniara Diagnostica LLC
- Product
- BIOPHEN Protein C 5 kit is for in vitro quantitative determination of Protein C activity on human citrated plasma. Protein C is a glycoprotein, vitamin K dependent, which inhibits coagulation. Protein C in plasma is measured following a specific activation with Protac (Agkistrodom Contortrix), an enzyme extracted from snake venom. Activated protein C hydrolysis the chromogenic substrate (SaPC-21) which release para-nitroaniline (pNA). The amount of pNA released (measured by absorbance at 405 nm) is directly proportional to the concentration of Protein C in the specimen.
- Product code
- GGP Test, Qualitative And Quantitative Factor Deficiency
- Reason for recall
- New warnings and precautionary statements on product labelling: Reagent 2 (R2) of BIOPHEN Protein C 5 contains the chemical substance - Cesium Chloride which changed classification and has become hazardous according to the CLP regulation ((EC) No 1272/2008) is suspected of damaging fertility
- Root cause tag
- labeling and UDI
- FDA root cause
- Other
- Action
- Aniara issued Medical Field Correction Letter (FSCA#29) prepared by HYPHEN BioMed and emailed to the customer by Aniara Diagnostica on 2/1/23. Letter states reason for recall, health risk and action to take: New labelling will be implemented on all new lots as follow: x A new pictogram added on reagent and box label: GHS08 x Signal Word: Warning x Hazard Statements: H361f - Suspected of damaging fertility x Precautionary Statements: o P201 : Obtain special instructions before use. o P280 : Wear protective gloves/protective clothing/eye protection/face protection. o P308 + P313 : IF exposed or concerned: Get medical advice/attention. o P405 : Store locked up. o P501 : Dispose of contents/container in consultation with your regional waste disposer x Hazardous ingredients : Cesium chloride The IFU, label and SDS will be updated. Please return completed and signed Form to your distributor.
- Quantity in commerce
- 2 units
- Distribution
- US Nationwide distribution in the state of NC.
- Product codes and lots
- UDI-DI: 366353700063BD Lot FA083327 (And all future lots)
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1227-2023 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.