Z-1196-2023Class II

MultiLab Series II REVO non-invasive vascular diagnostic device, Model Number 12950-0000-01

Unetixs Vascular, Inc. / initiated 2023-02-10 / Open, Classified

Unetixs Vascular, Inc. began a recall of MultiLab Series II REVO non-invasive vascular diagnostic device, Model Number 12950-0000-01 on . FDA classified it as Class II and gives the reason as "The current luer fittings and adaptors allow for the possibility of the air line to be misconnected to an intravenous line, which may cause an air embolism". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-1196-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Unetixs Vascular, Inc.
Product
MultiLab Series II REVO non-invasive vascular diagnostic device, Model Number 12950-0000-01
Product code
JAF Monitor, Ultrasonic, Nonfetal
Reason for recall
The current luer fittings and adaptors allow for the possibility of the air line to be misconnected to an intravenous line, which may cause an air embolism.
FDA root cause
Use error
Action
Urgent Medical Device Notification letters were sent to customers beginning 2/10/23. In order to physically prevent device misconnections from occurring, standardized connector designs for specific medical applications have been developed in accordance with ISO-80369-5. Unetixs Vascular has adopted this standard and ceased shipment of devices, hoses and blood pressure accessories equipped with luer fittings. All new devices will ship with bayonet connectors. These new connectors are designed to help reduce the risk for medical device misconnection. Action Items: 1. Immediately order one Bayonet Conversion Kit (BCK) per device via our website, www.unetixs.com/shop, to replace every air tubing connector on the device. 2. Complete the enclosed recall return response form. There is no need to return any product/accessories to Unetixs Vascular. Thank you for your attention to this matter. Unetixs Vascular is committed to providing high-quality products and partnering with you to ensure the safety of each patient. Please address any questions you may have with your local Sales Representative or our customer service department at 800-486-3849. If you are no longer in possession of this device, please ensure this notification gets to the current user. Please visit the following link for additional Examples of Medical Device Misconnections: https://www.fda.gov/medical-devices/medical-device-connectors/examples-medical-device-misconnections
Quantity in commerce
984 units
Distribution
Worldwide distribution - US Nationwide and Global distribution.
Product codes and lots
All Serial Numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1196-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1193-2023Class IIOngoingVoluntary: Firm initiated
Z-1194-2023Class IIOngoingVoluntary: Firm initiated
Z-1195-2023Class IIOngoingVoluntary: Firm initiated
Z-1196-2023Class IIOngoingVoluntary: Firm initiated