LATITUDE NXT Remote Patient Management System, LATITUDE NXT System Server Software Model 6460

Boston Scientific Corporation / initiated 2023-02-02 / Open, Classified / root cause: software

Boston Scientific Corporation began a recall of LATITUDE NXT Remote Patient Management System, LATITUDE NXT System Server Software Model 6460 on . FDA classified it as Class II and gives the reason as "Under specific circumstances, the U.S. product registration system did not send up enablement requests to U.S. LATITUDE Customer Support during an 8-month interval in 2022. For these patients, the HeartLogic Index and...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-1216-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Boston Scientific Corporation
Product
LATITUDE NXT Remote Patient Management System, LATITUDE NXT System Server Software Model 6460.
Product code
LWS Implantable Cardioverter Defibrillator (Non-Crt)
Reason for recall
Under specific circumstances, the U.S. product registration system did not send up enablement requests to U.S. LATITUDE Customer Support during an 8-month interval in 2022. For these patients, the HeartLogic Index and any associated Yellow Alerts are currently not visible in LATITUDE as may have been intended by the healthcare provider..
Root cause tag
software
FDA root cause
Software Manufacturing/Software Deployment
Action
The recalling firm issued letters to LATITUDE Healthcare Professionals via third-party Federal Express with signature at the account required on 2/2/2023. The letter informed the HCP that records indicate they may have U.S. patients who do not have HeartLogic enabled as the HCP may have intended in the Model 6460 U.S. LATITUDE NXT Patient Management System. The patients included within the enclosed list are scheduled to have HeartLogic enabled in LATITUDE ON 2/14/2023. The letter provides the clinical impact and recommendations. The recommendations include: (1) Distribute this information to appropriate HCPs within your center who manage HeartLogic alerts for the patients in the enclosed list; and (2) If you do NOT want HeartLogic enabled in LATITUDE for patients in the enclosed list, contact Boston Scientific LATITUDE Customer Support via [email protected] before 2/14/2023 and identify which patients should NOT be enabled. On 2/14/2023, BSC will correct the data on the Model 6460 LATITUDE Server for affected patients, unless a request not to enable is received by the firm prior to this date. Update: on 02/14/2023, Boston Scientific sent an additional letter, due to the original letter being lost in the mail. The revised letter advised customers to email LATITUDE Custoemr Support to have HeartLogic disabled.
Quantity in commerce
1,531 patient data records
Distribution
US Nationwide distribution to the healthcare facilities and healthcare providers are located in CA, CO, DE, FL, GA, ID, IL, IN, IA, KY, ME, MD, MA, NV, NH, NJ, NY, NC, OH, PA, RI, SC, TN TX, VA, WA, WV, and WI.
Product codes and lots
UDI-DI: 00802526613876

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1216-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1216-2023Class IIOngoingVoluntary: Firm initiated