In Vitro Diagnostic (IVD) Panel FilmArray GI Panel - REF RFIT-ASY-0116
BioFire Diagnostics, LLC began a recall of In Vitro Diagnostic (IVD) Panel FilmArray GI Panel - REF RFIT-ASY-0116 on . FDA classified it as Class II and gives the reason as "Due to temperature excursions with products (products being stored outside of the required temperature specifications) that may cause a reduction in performance. The temperature excursion was as a result of shipping...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1253-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- BioFire Diagnostics, LLC
- Product
- In Vitro Diagnostic (IVD) Panel FilmArray GI Panel - REF RFIT-ASY-0116
- Product code
- PCH Gastrointestinal Pathogen Panel Multiplex Nucleic Acid-Based Assay System
- Reason for recall
- Due to temperature excursions with products (products being stored outside of the required temperature specifications) that may cause a reduction in performance. The temperature excursion was as a result of shipping delays.
- Root cause tag
- manufacturing process
- FDA root cause
- Environmental control
- Action
- On 02/01/2023, the firm sent an "URGENT: MEDICAL DEVICE CORRECTION" Letter via email to customers informing them of a "Temperature Out of Range" issue with multiple products, as a result from Hurricane Ian (10/2022) Customers are Instructed to: Immediately examine their inventory for product identified in this recall. 1. Discontinue use and discard any remaining product in their possession that is subject to this recall. bioMerieux will replace the product at no charge in accordance with BIOFIRE s standard limited warranty. 2. Complete the accompanying Acknowledgment of Receipt Form and confirm the number of pouches scrapped (if any). Return the Acknowledgment of Receipt to bioMerieux so that bioMerieux may acknowledge their receipt of this notification. 3. If customers have further distributed this product, please identify any recipients, and notify them at once of this product recall. For questions or concerns, contact customer support department at [email protected] or via telephone by dialing +1.800.736.6354 and selecting option 5 for Product Technical Support
- Quantity in commerce
- 2 kits
- Distribution
- U.S. Nationwide distribution in the states of FL and GA.
- Product codes and lots
- Part Number: RFIT-ASY-0116 UDI: 00815381020109 Pouch Lot Number: 1674322
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1251-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1252-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1253-2023 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.