Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1228-2023Class II

Tri-Staple 2.0 Black Intelligent Reload, Product Number SIG60AXT

Medtronic, Inc. / initiated 2023-01-25 / Open, Classified

Covidien, LP began a recall of Tri-Staple 2.0 Black Intelligent Reload, Product Number SIG60AXT on . FDA classified it as Class II and gives the reason as "Affected lots have the potential for a broken sled vane, which may cause the reload to misfire leading to non-functional staple line closure, transecting tissue without forming staples, and tissue hang-up. These...". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-1228-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Medtronic, Inc.
Product
Tri-Staple 2.0 Black Intelligent Reload, Product Number SIG60AXT
Product code
GDW Staple, Implantable
Reason for recall
Affected lots have the potential for a broken sled vane, which may cause the reload to misfire leading to non-functional staple line closure, transecting tissue without forming staples, and tissue hang-up. These conditions may be associated to a delay to treatment, unspecified infection, hemorrhage/blood loss/bleeding, failure to anastomose, peritonitis, sepsis, pneumothorax, tissue trauma, or death.
FDA root cause
Process change control
Action
On January 25, 2023, the firm notified customers via Urgent Medical Device Recall letter. Customers were asked to take the following immediate actions: 1. Identify and quarantined all unused affected product. 2. Return unused affected product to Medtronic using the Customer Confirmation Form. 3. Forward the information to all who need to be aware within your organization, or to any organization where the affected product was transferred. There are no additional actions required for patients where a stapler in scope of this recall was used during a procedure. These patients should continue to be monitored in accordance with your medical facility s standard care protocols. If you have any questions regarding this communication, please contact your Medtronic Representative Customer Service at 800-962-9888, option 2.
Quantity in commerce
264 US; 5056 OUS
Distribution
Worldwide distribution - US Nationwide distribution in the states of Arkansas, California, Florida, Georgia, Louisiana, Massachusetts, New York, North Carolina, Ohio, Washington and the countries of Australia, Austria, Belgium, Canada, Canary Islands, Denmark, Finland, France, Germany, Ireland, Italy, Kuwait, Netherlands, New Caledonia, Norway, Poland, Portugal, Spain, Sweden, Switzerland, United Kingdom.
Product codes and lots
UDI-DI (GTIN): 20884521543598 Lots: N2D0002Y, N2D0004Y, N2D0195Y

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1228-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1228-2023Class IIOngoingVoluntary: Firm initiated