Z-1203-2023Class II

Phoenix mKDR Xpress, digital mobile diagnostic x-ray system

Sedecal S.A. / initiated 2023-02-06 / Open, Classified / root cause: design

SEDECAL SA began a recall of Phoenix mKDR Xpress, digital mobile diagnostic x-ray system on . FDA classified it as Class II and gives the reason as "Sedecal was informed of 3 incidents of erratic movements related to X-ray mobile systems". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-1203-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Sedecal S.A.
Product
Phoenix mKDR Xpress, digital mobile diagnostic x-ray system
Product code
IZL System, X-Ray, Mobile
Reason for recall
Sedecal was informed of 3 incidents of erratic movements related to X-ray mobile systems.
Root cause tag
design
FDA root cause
Device Design
Action
Sedecal issued an URGENT MEDICAL DEVICE RECALL notice to its US distributor on 02/06/2023 by email. The notice explained the issue and requested that distributor to warn its customers about the risk of potential erratic movement and provided a list of precautions until the software is installed. Questions may be directed to [email protected].
Quantity in commerce
7 units
Distribution
US, Argentina, Panama
Product codes and lots
UDI/DI 08436046003163, Serial Numbers: PH00202, PH00203, PH00204, PH00205, PH00206, PH00207, PH00208