Z-1190-2023Class I

BALLARD* ACCESS Closed Suction System for Neonates/Pediatrics, Elbow Manifold

Halyard Health / initiated 2023-02-03 / Terminated / root cause: design

Avanos Medical, Inc. began a recall of BALLARD* ACCESS Closed Suction System for Neonates/Pediatrics, Elbow Manifold on . FDA classified it as Class I and gives the reason as "Potential for product manifold to crack during use on patients, possibly causing leaks in the respiratory circuit and can lead to the patient receiving inadequate ventilation". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1190-2023
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Halyard Health
Product
BALLARD* ACCESS Closed Suction System for Neonates/Pediatrics, Elbow Manifold.
Product code
BSY Catheters, Suction, Tracheobronchial
Reason for recall
Potential for product manifold to crack during use on patients, possibly causing leaks in the respiratory circuit and can lead to the patient receiving inadequate ventilation.
Root cause tag
design
FDA root cause
Component design/selection
Action
Avanos hand-delivered the recall notification to the single consignee on 02/03/2023. The letter instructed the customer to discontinue use of and quarantine their stock of all lots of affected units and return them. Additionally, the customer was instructed to complete and return the acknowledgement form.
Quantity in commerce
220 units
Distribution
US Nationwide distribution in the state of Oklahoma.
Product codes and lots
All lots are affected. Stock Code (Lot Number): B50511 (20098277), B50611 (20098278), B50711 (20098279), B50811 (20097177, 20097177R1).

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1190-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1189-2023Class ITerminatedVoluntary: Firm initiated
Z-1190-2023Class ITerminatedVoluntary: Firm initiated