Z-1191-2023Class II

stryker Hoffmann LRF, Transport Strut, Model Number 4933-0-400; Orthopedic external fixation device intended...

Stryker Corp. / initiated 2023-02-07 / Open, Classified

Stryker GmbH began a recall of stryker Hoffmann LRF, Transport Strut, Model Number 4933-0-400; Orthopedic external fixation device intended to keep bone fragments/segments in a correct... on . FDA classified it as Class II and gives the reason as "Stryker has identified a nonconformance in specific lots of Hoffmann LRF Bone Transport Struts. Specifically, the thread pitch is oversized at the distal end of the threaded rod, and as a result the distal ring fixation...". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-1191-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Stryker Corp.
Product
stryker Hoffmann LRF, Transport Strut, Model Number 4933-0-400; Orthopedic external fixation device intended to keep bone fragments/segments in a correct anatomical position until reliable bone consolidation is achieved
Product code
KTT Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component
Reason for recall
Stryker has identified a nonconformance in specific lots of Hoffmann LRF Bone Transport Struts. Specifically, the thread pitch is oversized at the distal end of the threaded rod, and as a result the distal ring fixation cannot be locked between 0- 50mm of the threaded rod.
FDA root cause
Process change control
Action
The firm issue an URGENT MEDICAL DEVICE RECALLL its consignees by letter on 02/07/2023. The notice explained the product and the issue and requested the following actions be taken: 1. Check inventory for affected product and remove from use 2, Sign and return the enclosed Business Reply Form by email to [email protected] to confirm receipt of this notification/documenting product disposition. Stryker will will contact you to arrange for the return of the affected device(s) 3. Spread awareness of this communication internally until all required actions have been completed within your facility. 4. Inform Stryker if any of the subject devices have been distributed to other organizations. If so, provide contact details so that Stryker can inform the recipients appropriately.
Quantity in commerce
488 units
Distribution
US Nationwide distribution.
Product codes and lots
Serial numbers: D32158, D32160, D32161, D32162, D32163, D32164, D34343, D34344, D40690, D41629, D41630, D41631, D44713, G16331, G16332, G16333, G26278, G29759, G29760, G32730, G32731, G32735, G45119, H30398, H30399, H66392

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1191-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1191-2023Class IIOngoingVoluntary: Firm initiated