Z-1245-2023Class II

Fusion Bioline Vascular Graft, Part number M00201503046B0

Maquet Cardiovascular LLC / initiated 2023-01-30 / Open, Classified / root cause: manufacturing process

Maquet Cardiovascular, LLC began a recall of Fusion Bioline Vascular Graft, Part number M00201503046B0 on . FDA classified it as Class II and gives the reason as "One (1) Fusion Bioline Vascular graft from lot 25162546, which failed guideline straightness testing, was intermingled with lot 25162549 and distributed to customers. Use of the affected graft may result in reduction of...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1245-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Maquet Cardiovascular LLC
Product
Fusion Bioline Vascular Graft, Part number M00201503046B0
Product code
DSY Prosthesis, Vascular Graft, Of 6mm And Greater Diameter
Reason for recall
One (1) Fusion Bioline Vascular graft from lot 25162546, which failed guideline straightness testing, was intermingled with lot 25162549 and distributed to customers. Use of the affected graft may result in reduction of blood flow or unwanted tension causing bleeding and the formation of a pseudo-aneurysm.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On January 30, 2023, the firm notified customers via Urgent Medical Device Removal letters. Customers are asked to examine your inventory immediately for lot 25162546. Please note, the Lot number appears on the device shelf boxes only. If you have the affected lot, please stop using and remove the device from areas of use. You are entitled to a replacement at no cost to your facility. You will receive replacement upon your acknowledgement that you have product for return. Distributors who have shipped this product should notify their customers. Please contact Getinge Customer Service at 888-943-8872 between the hours of 8:00 a.m. and 6:00 p.m. Eastern Standard Time to request a return authorization (RMA) and shipping instructions to return any affected product.
Quantity in commerce
1 (US)
Distribution
US Nationwide distribution in the states of FL, IN, LA, MI, MO, NC, NV, NY, PA, TN, TX, VA, WA.
Product codes and lots
UDI-DI 00384409005942 Lot 25162546

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1245-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1245-2023Class IIOngoingVoluntary: Firm initiated