DeRoyal SafeLiner Suction Canister, 1000cc, REF PHESL-1000B
DeRoyal Industries Inc began a recall of DeRoyal SafeLiner Suction Canister, 1000cc, REF PHESL-1000B on . FDA classified it as Class II and gives the reason as "The reason for the voluntary recall is due to shrinkage of the canister lid. This shrinkage causes the lid not to contact the outer shell to create a seal and thus, it does not create and hold vacuum". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1283-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- DeRoyal Industries, Inc.
- Product
- DeRoyal SafeLiner Suction Canister, 1000cc, REF PHESL-1000B
- Product code
- KDQ Bottle, Collection, Vacuum
- Reason for recall
- The reason for the voluntary recall is due to shrinkage of the canister lid. This shrinkage causes the lid not to contact the outer shell to create a seal and thus, it does not create and hold vacuum.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- The firm issued an URGENT DEROYAL RECALL NOTICE its consignees by email and letter beginning on 02/09/2023. The notice explained the problem and risk, and requested the following actions be taken: Identify and quarantine any affected product found in inventory. If the product has been further distributed, please notify then of this recall. Firm is seeking product return. Questions about receiving credit for return merchandise, contact Jennifer Marsh at [email protected] or by phone 865-362-4203 or email us at [email protected]. If you have questions regarding this recall please contact us at 865-362-6465 or email us at [email protected].
- Quantity in commerce
- 27,200 devices
- Distribution
- US Nationwide distribution in the states of Arizona, Texas, Virginia, Pennsylvania, California, Missouri, North Carolina, Illinois, Indiana, Ohio, Arkansas.
- Product codes and lots
- GTIN 50749756377046, Lot 58447585 exp. 12/22/2027
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1283-2023 | Class II | Ongoing | Voluntary: Firm initiated |