Z-1292-2023Class II

Cirq Arm System 2.0

Brainlab, Inc. / initiated 2023-03-02 / Open, Classified / root cause: manufacturing process

Brainlab AG began a recall of Cirq Arm System 2.0 on . FDA classified it as Class II and gives the reason as "Risk of mechanical instability due to potential manufacturing error of the device". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1292-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Brainlab, Inc.
Product
Cirq Arm System 2.0
Product code
OLO Orthopedic Stereotaxic Instrument
Reason for recall
Risk of mechanical instability due to potential manufacturing error of the device
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The firm issued a URGENT FIELD SAFETY NOTICE / PRODUCT NOTIFICATION to their consignees by email on 03/02/2023. The notice explained the issue, risk of use, and requested the following actions be taken: "1. Do not use Cirq Arm System 2.0 for clinical procedures. 2. Remove Cirq Arm System 2.0 from the OR table (as applicable), following the description given in the applicable User Guide. When lifting Cirq Arm System 2.0, ensure you carry the device only below joint 4 (see picture 1) to prevent user injury from falling parts. 3. Store the device in the corresponding transport and storage case delivered with it and ensure you mark the equipment appropriately to prevent it from being used inadvertently." Brainlab is seeking return of the devices for inspection and repair (if applicable).
Quantity in commerce
1 unit
Distribution
Texas
Product codes and lots
56500 CIRQ ARM SYSTEM 2.0; GTIN 04056481143961; All available Serial numbers/lot numbers are affected.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1292-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1292-2023Class IIOngoingVoluntary: Firm initiated