Moberg Component Neuromonitoring System (CNS Monitor)-Intended as a short-term neurological monitoring system...

Moberg Research, Inc. / initiated 2023-02-20 / Open, Classified

Moberg Research, Inc. began a recall of Moberg Component Neuromonitoring System (CNS Monitor)-Intended as a short-term neurological monitoring system that collects, displays, and stores multiple... on . FDA classified it as Class II and gives the reason as "Battery leakage can cause corrosion to the metal enclosure near the electrical interface. Prolonged contact to corrosive materials could cause injury to skin. DOC". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-1338-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Moberg Research, Inc.
Product
Moberg Component Neuromonitoring System (CNS Monitor)-Intended as a short-term neurological monitoring system that collects, displays, and stores multiple physiological measurements Model: CNS-350
Product code
OMA Amplitude-Integrated Electroencephalograph
Reason for recall
Battery leakage can cause corrosion to the metal enclosure near the electrical interface. Prolonged contact to corrosive materials could cause injury to skin. DOC
FDA root cause
Under Investigation by firm
Action
Natus issued Letters Urgent Medical Device Recall Letter Component Neuromonitoring System (Moberg CNS Monitor Model CNS-350) via first class beginning Feb 20, 2023. Letter states reason for recall, health risk and action to take: Ensure that the CNS is either shut down or plugged into a power outlet. Perform the next step while using proper personal protective equipment (PPE), including gloves. On the top of the CNS-350 device, open each of the three (3) battery ports and remove any backup battery. 2) Inspect the battery and the battery port for any white, brown, or green/blue buildup, or for any corrosion of the metal contact locations. a. Dispose of the battery at your local electronic e-waste facility. Note: The CNS-350 device may continue to be used (while plugged into a power outlet) with the backup battery removed. b. If battery leakage or corrosion is suspected, cease use of the CNS-350 device. New replacement battery will be shipped to you. The Moberg Technical Service department will be in contact with you in relation to the provision of the replacement product. For new battery replacement, installation instructions will be provided with the battery.
Quantity in commerce
89 units
Distribution
Worldwide distribution - US Nationwide and the countries of France, Germany, Italy, Sweden.
Product codes and lots
UDI Number: B485CNS3501 Serial Numbers: A-CNS-22502 A-CNS-22506 A-CNS-22507 A-CNS-22508 A-CNS-22509 A-CNS-22510 A-CNS-22511 A-CNS-22512 A-CNS-22513 A-CNS-22525 A-CNS-22526 A-CNS-22527 A-CNS-22528 A-CNS-22529 A-CNS-22530 A-CNS-22534 A-CNS-22535 A-CNS-22536 A-CNS-22537 A-CNS-22541 A-CNS-22542 A-CNS-22543 A-CNS-22544 A-CNS-22545 A-CNS-22546 A-CNS-22547 A-CNS-22548 A-CNS-22549 A-CNS-22550 A-CNS-22551 A-CNS-22552 A-CNS-22553 A-CNS-22554 A-CNS-22555 A-CNS-22556 A-CNS-22557 A-CNS-22558 A-CNS-22559 A-CNS-22560 A-CNS-22561 A-CNS-22562 A-CNS-22564 A-CNS-22565 A-CNS-22566 A-CNS-22567 A-CNS-22568 A-CNS-22569 A-CNS-22570 A-CNS-22571 A-CNS-22572 A-CNS-22573 A-CNS-22574 A-CNS-22575 A-CNS-22577 A-CNS-22578 A-CNS-22584 A-CNS-22585 A-CNS-22586 A-CNS-22587 A-CNS-22588 A-CNS-22589 A-CNS-22590 A-CNS-22591

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1338-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1338-2023Class IIOngoingVoluntary: Firm initiated