INFINITY Occipitocervical Upper Thoracic System, MULTI AXIAL SCREW, Size: 4.0mm x 26mm, REF 3604026; for...
Medtronic Sofamor Danek USA Inc began a recall of INFINITY Occipitocervical Upper Thoracic System, MULTI AXIAL SCREW, Size: 4.0mm x 26mm, REF 3604026; for immobilization and stabilization of spinal segment as... on . FDA classified it as Class II and gives the reason as "Incorrect type of titanium was used which reduces the gripping strength of the screw head of 5 batches of Infinity Multi-Axial Screws". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-1412-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Medtronic Sofamor Danek USA Inc
- Product
- INFINITY Occipitocervical Upper Thoracic System, MULTI AXIAL SCREW, Size: 4.0mm x 26mm, REF 3604026; for immobilization and stabilization of spinal segment as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical unction, the cervical spine (C1 to C7), and the thoracic spine from T1-T3.
- Product code
- KWP Appliance, Fixation, Spinal Interlaminal
- Reason for recall
- Incorrect type of titanium was used which reduces the gripping strength of the screw head of 5 batches of Infinity Multi-Axial Screws
- FDA root cause
- Under Investigation by firm
- Action
- The firm issued an URGENT MEDICAL DEVICE RECALL NOTICE to its consignees on 02/16/2023 by UPS 2nd day mail. The letter explained the issue and the risk and requested the following actions be taken: Immediately identify, segregate, and quarantine affected products within your inventory. Pass this notice to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. Contact Medtronic to return affected product: 1-800-933-2635 to receive an return authorization (RGA) and coordinate inventory replacement/credit. Reference FA1315. The letter also includes specific patient recommendations.
- Quantity in commerce
- 90 units
- Distribution
- Worldwide distribution - US Nationwide and the country of Canada.
- Product codes and lots
- GTIN 00643169554696, Lot/Batch numbers: H5806345
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1409-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1410-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1411-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1412-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1413-2023 | Class II | Ongoing | Voluntary: Firm initiated |