Z-1416-2023Class II
SmartPerfusion, release R1.0, R1.1, R1.1.1, R1.1.5, R1.1.5.1 and R1.1.6
Philips Medical Systems Nederland B.V. / initiated 2023-02-20 / Open, Classified / root cause: software
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. began a recall of SmartPerfusion, release R1.0, R1.1, R1.1.1, R1.1.5, R1.1.5.1 and R1.1.6 on . FDA classified it as Class II and gives the reason as "There are technical issues related to signal generation and processing, which can lead to inaccurate presentations". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1416-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Philips Medical Systems Nederland B.V.
- Product
- SmartPerfusion, release R1.0, R1.1, R1.1.1, R1.1.5, R1.1.5.1 and R1.1.6
- Product code
- OWB Interventional Fluoroscopic X-Ray System
- Reason for recall
- There are technical issues related to signal generation and processing, which can lead to inaccurate presentations.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- A recall notification letter was issued February 20, 2023. The Urgent Medical Device Correction letter includes the following instructions for the Customer: " Stop using SmartPerfusion and 2D Perfusion. NOTE: This message does not impact (other parts of) the Philips Allura or Azurion systems or other tools within the Interventional Workspot. The Philips Allura and Azurion systems as well as other Interventional Workspot tools may continue to be used. " Place the Urgent Medical Device Correction Letter with the documentation of the system. " Circulate the notice to all users so that they are aware of the product issue. " Return the reply form attached to the Urgent Medical Device Correction letter to Philips to confirm that the users of the system have reviewed and understood it.
- Quantity in commerce
- 1461 units (236 US, 1225 OUS)
- Distribution
- Worldwide - US Nationwide distribution.
- Product codes and lots
- UDI: (01)00884838088566 SmartPerfusion with Allura Xper or Azurion model number: 722003, 722006, 722008, 722010, 722012, 722013, 722020, 722023, 722026, 722027 722028, 722029, 722035, 722038, 722039, 722058, 722059, 722064, 722067 722068, 722078, 722079, 722223, 722224, 722225, 722226, 722227 and 722228. Software license key: 459801530271 sw key SmartPerfusion
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1416-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1417-2023 | Class II | Ongoing | Voluntary: Firm initiated |