Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1294-2023Class II

CoCr Femoral Head, XS, 38/-8, Taper 12/14-Intended to be used as a modular head component for articulation in...

Zimmer GmbH / initiated 2023-02-27 / Open, Classified / root cause: labeling and UDI

Zimmer, Inc. began a recall of CoCr Femoral Head, XS, 38/-8, Taper 12/14-Intended to be used as a modular head component for articulation in total hip arthroplasty. Item Number: 01.01012.384 on . FDA classified it as Class II and gives the reason as "Update the compatibility matrix as referred to in the Instructions for Use (IFU) for the CoCr Femoral Head XS- removes the compatibility with the Epsilon Durasul Constrained Acetabular Liners from the matrix due to the...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1294-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Zimmer GmbH
Product
CoCr Femoral Head, XS, 38/-8, Taper 12/14-Intended to be used as a modular head component for articulation in total hip arthroplasty. Item Number: 01.01012.384
Product code
LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
Reason for recall
Update the compatibility matrix as referred to in the Instructions for Use (IFU) for the CoCr Femoral Head XS- removes the compatibility with the Epsilon Durasul Constrained Acetabular Liners from the matrix due to the range of motion in flexion/extension being less than 100 degrees as recommended per an internationally recognized standard (ISO 21535:2009
Root cause tag
labeling and UDI
FDA root cause
Under Investigation by firm
Action
Zimmer Biomet issued Urgent Medical Device Correction letter on 2/27/23 via Fed'X .Letter to Distributors, Doctors, Risk Managers and states reason for recall, health risk and action to take: Hospital Responsibilities: 1. Review this field action notice and ensure that affected personnel are aware of the contents. 2. Complete Attachment 1 Certificate of Acknowledgement Form and send to [email protected]. This form must be returned even if you do not have affected products at your facility. 3. Retain a copy of the Attachment 1 Certificate of Acknowledgement Form with your field action notice records in the event of a compliance audit of your facility s documentation. 4. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to [email protected]. Surgeon Responsibilities: 1. Review this field action notice for awareness of the contents. 2. There are no specific patient monitoring instructions related to this notice that are recommended beyond your existing follow-up schedule. 3. Complete Attachment 1 Certificate of Acknowledgement Form and send to [email protected] . 4. Retain a copy of the Attachment 1 Certificate of Acknowledgement Form with your field action notice records in the event of a compliance audit of your facility s documentation. 5. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an
Quantity in commerce
1,168 untits
Distribution
AL AZ CO FL LA MI OH OK TX Foreign: Canada
Product codes and lots
GTIN Number: 00889024283268 All lots

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1294-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1294-2023Class IIOngoingVoluntary: Firm initiated