SafeLight Fiber Optic Cable: Clear, 5MM X 10FT 3.05M, REF: 0233-050-100; Gray, 5MM X 10FT 3.05M, REF...
Stryker Corporation began a recall of SafeLight Fiber Optic Cable: Clear, 5MM X 10FT 3.05M, REF: 0233-050-100; Gray, 5MM X 10FT 3.05M, REF: 0233-050-200 on . FDA classified it as Class II and gives the reason as "Fiber optic cables assembled with not enough epoxy on the proximal end, creating light output issues, which could cause a clinical delay or procedure being converted to open". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1474-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Stryker Corp.
- Product
- SafeLight Fiber Optic Cable: Clear, 5MM X 10FT 3.05M, REF: 0233-050-100; Gray, 5MM X 10FT 3.05M, REF: 0233-050-200
- Product code
- FCW Light Source, Fiberoptic, Routine
- Reason for recall
- Fiber optic cables assembled with not enough epoxy on the proximal end, creating light output issues, which could cause a clinical delay or procedure being converted to open.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- On 3/2/23 recall notices were mailed to customers who were asked to do the following: 1. Inform individuals within your organization who need to be aware of this action. 2. Identify, remove, and quarantine affected devices. 3. If affected product is found, call customer service at 1-800-624-4422 (Option 3) or email [email protected] to arrange for product return. 4. Complete and return the Business Reply Form Customers with questions can contact [email protected]
- Quantity in commerce
- 5,228
- Distribution
- Worldwide - US Nationwide distribution including in the states of PA, MS, IL, OH, CA, TX, MO, IN, KS, OK, MT, MA, NE, AZ, NV, NJ, IA, WV, MD, KY, FL, NM, NY, CO, WA, NC, VA, LA, ND, WY, HI SC, UT, OR, SD, AR, TN, AK, CT, GA, ID, VT, NH, MN, WI, DE, RI, AL, MI and the countries of NL, PE, AT, CA, TW, PH, SG, KR, CN.
- Product codes and lots
- Ref/UDI-DI/Lot: 0233-050-100/07613327051650/159190, 159191, 159192, 159194, 159195, 159196, 159197, 159198, 159199, 159200, 159201, 159202, 159203, 159204, 159205, 159206, 159207, 159208, 159209, 159210, 159211, 159212, 159213, 159214, 159215, 159216, 159217, 159218, 164764, 164765, 164766, 164767, 164768, 164769, 164770, 164771, 164772, 164773, 164774, 164775, 164776, 164777, 164778, 164779, 164780, 164781, 164782, 164783, 164784, 164785, 164786, 164787, 164788, 164789, 164790, 164791, 164792, 164793, 164794, 164795, 164796, 164797, 164798, 164799, 164800, 164801, 164802, 164803, 164804, 164805, 164806, 164807, 164808, 164809, 164810, 164811, 164812, 164813, 164814, 164815, 164816, 164817, 164818, 164819, 164820, 164821, 164822, 168940, 168941, 168942, 168943, 168944, 168945, 168946, 168947, 168948, 168949, 168950, 168951, 168952, 168953, 168954, 168955, 168956, 168957, 168958, 168959, 168960, 168961, 168962, 168963, 168964, 168965, 168966, 168967, 168968, 168969, 168970, 168971, 168972, 168973, 168974, 168975, 168976, 168977, 168978, 168979, 168980, 168981, 168987; 0233-050-200/07613327051643/158865, 158868, 158869, 158872, 158874, 158875, 158878, 158884, 165446, 165448, 165451, 165457, 165458, 165459, 165461, 165462, 165464, 168909, 168910, 168911, 168912, 168913, 168914, 168915, 168916, 168917, 168918, 168919, 168920, 168921, 168923, 168924, 168925, 168926, 168927, 168928
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1474-2023 | Class II | Ongoing | Voluntary: Firm initiated |