EMPOWR Porous Knee Patella Over Drill, REF: 802-05-104; EMPOWR Knee Patella Drill, REF: 802-05-108
Encore Medical, LP began a recall of EMPOWR Porous Knee Patella Over Drill, REF: 802-05-104; EMPOWR Knee Patella Drill, REF: 802-05-108 on . FDA classified it as Class II and gives the reason as "Patella drills are marked with incorrect diameter size of 5.6mm, when they should be 5.4mm. Porous patella overdrills are marked with incorrect diameter size of 5.4 mm, when they should be 5.6 mm. If the incorrect drill...". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-1384-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Encore Medical, LP
- Product
- EMPOWR Porous Knee Patella Over Drill, REF: 802-05-104; EMPOWR Knee Patella Drill, REF: 802-05-108
- Product code
- MBH Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer
- Reason for recall
- Patella drills are marked with incorrect diameter size of 5.6mm, when they should be 5.4mm. Porous patella overdrills are marked with incorrect diameter size of 5.4 mm, when they should be 5.6 mm. If the incorrect drill is selected during surgery, there is risk of the implant loosening, implant not properly fitting, and/or fractured bone.
- FDA root cause
- Process change control
- Action
- On 2/28/23, field safety notices were emailed to surgical sales agents who were asked to document the quantity of affected devices complete and return the acknowledgement form. Customer service contact, [email protected], was provided to report failures or to request replacements. Customers with questions can phone (914) 450-3931, or email [email protected].
- Quantity in commerce
- 29
- Distribution
- US: TX, IN, MD, CA, KS, AL, HI, WA
- Product codes and lots
- REF/UDI-DI/Lot: 802-05-104/00190446238843/388620L01; 802-05-108/00190446365907/388620L03
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1384-2023 | Class II | Ongoing | Voluntary: Firm initiated |