Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1385-2023Class II

stryker MAKO RESTORIS MCK BASEPLATE TRIAL (SIZE 5 RM/LL), REF 170615, intended for use in a Partial Knee...

Howmedica Osteonics Corp / initiated 2023-03-10 / Open, Classified

Howmedica Osteonics Corp. began a recall of stryker MAKO RESTORIS MCK BASEPLATE TRIAL (SIZE 5 RM/LL), REF 170615, intended for use in a Partial Knee Arthroplasty surgical procedure on . FDA classified it as Class II and gives the reason as "Ccertain lots of the Restoris MultiCompartmental Knee (MCK) Tibial Baseplate Trials contain product with incorrect size laser marking". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-1385-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Howmedica Osteonics Corp
Product
stryker MAKO RESTORIS MCK BASEPLATE TRIAL (SIZE 5 RM/LL), REF 170615, intended for use in a Partial Knee Arthroplasty surgical procedure
Product code
HSX Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer
Reason for recall
Ccertain lots of the Restoris MultiCompartmental Knee (MCK) Tibial Baseplate Trials contain product with incorrect size laser marking
FDA root cause
Process change control
Action
Stryker issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 03/10/2023 by letter. The notice explained the issue and the hazard and requested the following actions be taken: 1. Please inform users of this Urgent Medical Device Recall and forward this notice to all individuals who need to be made aware. 2. Immediately check all stock areas and/or operating room storage to determine if any devices from the affected product list are at your facility. 3. Quarantine and discontinue use of the affected products in Table 1. 4. Complete and sign the enclosed Urgent Medical Device Recall Business Reply Form and email to [email protected]/ fax (888) 266-7910. 5. Please contact your Local Sales Office or your Stryker Sales Representative directly for product replacement and inventory questions. 6. Please return ALL affected product to: Stryker Orthopaedics/PFA Product Return. Stryker issued an URGENT MEDICAL DEVICE RECALL UPDATE notice to its consignees on 04/12/2023. The letter contained the same information, but added a section titled Recommendations for patients already treated with an affected device, and advised patients treated with an affected product should continue to be followed per the normal protocol established by his or her surgeon(s).
Quantity in commerce
52 units
Distribution
U.S., Netherlands, Japan, China, India, Korea, and Australia.
Product codes and lots
GTIN 00848486003746, Lot Number 26270421

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1385-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1385-2023Class IIOngoingVoluntary: Firm initiated
Z-1386-2023Class IIOngoingVoluntary: Firm initiated