Z-1385-2023Class II
stryker MAKO RESTORIS MCK BASEPLATE TRIAL (SIZE 5 RM/LL), REF 170615, intended for use in a Partial Knee...
Howmedica Osteonics Corp / initiated 2023-03-10 / Open, Classified
Howmedica Osteonics Corp. began a recall of stryker MAKO RESTORIS MCK BASEPLATE TRIAL (SIZE 5 RM/LL), REF 170615, intended for use in a Partial Knee Arthroplasty surgical procedure on . FDA classified it as Class II and gives the reason as "Ccertain lots of the Restoris MultiCompartmental Knee (MCK) Tibial Baseplate Trials contain product with incorrect size laser marking". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-1385-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Howmedica Osteonics Corp
- Product
- stryker MAKO RESTORIS MCK BASEPLATE TRIAL (SIZE 5 RM/LL), REF 170615, intended for use in a Partial Knee Arthroplasty surgical procedure
- Product code
- HSX Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer
- Reason for recall
- Ccertain lots of the Restoris MultiCompartmental Knee (MCK) Tibial Baseplate Trials contain product with incorrect size laser marking
- FDA root cause
- Process change control
- Action
- Stryker issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 03/10/2023 by letter. The notice explained the issue and the hazard and requested the following actions be taken: 1. Please inform users of this Urgent Medical Device Recall and forward this notice to all individuals who need to be made aware. 2. Immediately check all stock areas and/or operating room storage to determine if any devices from the affected product list are at your facility. 3. Quarantine and discontinue use of the affected products in Table 1. 4. Complete and sign the enclosed Urgent Medical Device Recall Business Reply Form and email to [email protected]/ fax (888) 266-7910. 5. Please contact your Local Sales Office or your Stryker Sales Representative directly for product replacement and inventory questions. 6. Please return ALL affected product to: Stryker Orthopaedics/PFA Product Return. Stryker issued an URGENT MEDICAL DEVICE RECALL UPDATE notice to its consignees on 04/12/2023. The letter contained the same information, but added a section titled Recommendations for patients already treated with an affected device, and advised patients treated with an affected product should continue to be followed per the normal protocol established by his or her surgeon(s).
- Quantity in commerce
- 52 units
- Distribution
- U.S., Netherlands, Japan, China, India, Korea, and Australia.
- Product codes and lots
- GTIN 00848486003746, Lot Number 26270421
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1385-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1386-2023 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY