Medtronic SenSight Extension Tunneler Kit, REF B31030; The Extension Tunneler Kit is used to assist in the...
Medtronic Neuromodulation began a recall of Medtronic SenSight Extension Tunneler Kit, REF B31030; The Extension Tunneler Kit is used to assist in the tunneling and passage of extensions for deep brain... on . FDA classified it as Class II and gives the reason as "Medtronic has received reports of SenSight Extension Tunneler kits that contain dual carriers with have been machined on only one side. As a result, the dual carrier cannot be used to pass two extensions simultaneously". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1408-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Medtronic Neuromodulation
- Product
- Medtronic SenSight Extension Tunneler Kit, REF B31030; The Extension Tunneler Kit is used to assist in the tunneling and passage of extensions for deep brain stimulation
- Product code
- MHY Stimulator, Electrical, Implanted, For Parkinsonian Tremor
- Reason for recall
- Medtronic has received reports of SenSight Extension Tunneler kits that contain dual carriers with have been machined on only one side. As a result, the dual carrier cannot be used to pass two extensions simultaneously.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Medtronic issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 03/17/2023 via UPS 2-day delivery. The notice explained the issue and the hazard and requested the following actions be taken: Quarantine all unused product from the affected lots and return them as described on the Customer Confirmation Form. Pass the notice to all those who need to be aware within your organization or to any organization where the potentially affected product has been transferred or distributed.
- Quantity in commerce
- 1414 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canary Islands, Czech Republic, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Japan, Luxembourg, Netherlands, New Zealand, Portugal, Slovakia, Spain, Sweden, Switzerland, United Kingdom.
- Product codes and lots
- UDI/DI00763000426361, Serial Numbers: HG56BR6, HG56C4R, HG56CJAH01, HG56CLE, HG586N7, HG586UM, HG58704, HG58VM0, HG58VQH, HG58VQHH01, HG58VY5, HG58W7H, HG58WWC, HG58WYB, HG58WYBH01, HG58X0E, HG597BH, HG597DH, HG597M8, HG59ABH, HG59AFC, HG59AV6, HG59RY0, HG5A9GT, HG5A9VY, HG5A9VZ, HG5AQH0, HG5AQQE, HG5BQ5Y, HG5BQLR, HG5BQPJ, HG5BTW7, HG5BTZX, HG5BU5W, HG5BU77, HG5EPSB, HG5EQ0C, HG5EQ2A, HG5EQ69, HG5F31N, HG5F338, HG5F3AZ, HG5F3EH, HG5F3G6, HG5F3K5, HG5F3NF, HG5FP0Y, HG5FP3B, HG5FP70, HG5FP70H01, HG5FPBE, HG5HQBM, HG5HQF8, HG5LJTL, HG5LJZQ, HG5LK2J, HG5LK4H, HG5LK5Y, HG5LK7C, HG5LK8P, HG5LK92, HG5LKAL, HG5LKC9, HG5LKEL, HG5LKGT, HG5LR02, HG5LR33, HG5LTWA, HG5LTWP, HG5LTXU, HG5LTYY, HG5LU12; UDI/DI 00763000486518, Serial Numbers: HG59ABH, HG59RY0, HG59SJ4, HG59SNC, HG5A9H3, HG5A9NB, HG5A9VY, HG5BQ5Y, HG5BQAW, HG5BQLR.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1408-2023 | Class II | Ongoing | Voluntary: Firm initiated |