Z-1426-2023Class II

JULIET Ti LL, LATERAL LUMBAR TI-LIFE PLATE SMALL, REF: JLT-PL 02 08-S, JLT-PL 02 10-S, JLT-PL 02 12-S, JLT-PL...

Spineart SA / initiated 2023-03-16 / Open, Classified

SPINEART SA began a recall of JULIET Ti LL, LATERAL LUMBAR TI-LIFE PLATE SMALL, REF: JLT-PL 02 08-S, JLT-PL 02 10-S, JLT-PL 02 12-S, JLT-PL 02 14-S, JLT-PL 02 16-S; JULIET Ti LL, LATERAL... on . FDA classified it as Class II and gives the reason as "JULIET Lateral Lumbar Ti-Life Plate, a component of the JULIET Ti Lateral Lumbar Cage, may disassemble during the preparation stage before implantation, or during surgery when removing the implant holder, which could...". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-1426-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Spineart SA
Product
JULIET Ti LL, LATERAL LUMBAR TI-LIFE PLATE SMALL, REF: JLT-PL 02 08-S, JLT-PL 02 10-S, JLT-PL 02 12-S, JLT-PL 02 14-S, JLT-PL 02 16-S; JULIET Ti LL, LATERAL LUMBAR TI-LIFE PLATE LARGE, REF: JLT-PL 04 08-S, JLT-PL 04 10-S, JLT-PL 04 12-S, JLT-PL 04 14-S, JLT-PL 04 16-S, JLT-PL 04 18-S
Product code
OVD Intervertebral Fusion Device With Integrated Fixation, Lumbar
Reason for recall
JULIET Lateral Lumbar Ti-Life Plate, a component of the JULIET Ti Lateral Lumbar Cage, may disassemble during the preparation stage before implantation, or during surgery when removing the implant holder, which could cause surgical delay.
FDA root cause
Process change control
Action
On 3/16/23, recall notices were distributed to customers who were told the following: Plate replacement: If the issue occurs, the surgeon should proceed as follows: 1- Unscrew the Plate screw of the cage with the screwdriver, 2- Scrap the current plate assembly, 3- Use a new set of pre-assembled plate/screw (a minimum of 2 sets of pre-assembled plates/screws per size are part of the device listing). The cages are to be used with the corresponding plate or higher plate which means that a plate with the same height than the cage can be used but also a plate with higher height. 4- In the case of 0-degree and 8-degree cages, a pre-assembled plate/screw is not mandatory as described in the surgical technique Acknowledge receipt of this letter by emailing to [email protected] Discontinue use and return any product on hand. Customers with questions can contact firm at 949-313-6400 Monday through Friday, 8:00am to 5:00pm, Pacific, or email at [email protected]
Quantity in commerce
544
Distribution
US Nationwide distribution in the states of NH, TX, CA, VA, IL, RI, NC, AZ.
Product codes and lots
REF/UDI-DI/Lot: JLT-PL 02 08-S/07640178987388/4-8826, 5-4142, 5-5377, 5-0643, 5-2431; JLT-PL 02 10-S/07640178987395/6-5488, 6-6556, 5-0644, 6-4734, 6-4735, 5-5378, 6-3211; JLT-PL 02 12-S/07640178987401/6-5489, 6-6558, 6-7645, 6-8734, 6-4732, 4-8828, 5-5379, 6-4733; JLT-PL 02 14-S/07640178987418/5-8439, 4-8829, 6-0714, 5-5380; JLT-PL 02 16-S/07640178987425/4-8830, 5-4145, 6-3217, 6-0726, 6-5490; JLT-PL 04 08-S/07640178987456/4-8845, 5-4146, 6-0725; JLT-PL 04 10-S/07640178987463/4-8831, 5-4147, 6-0722; JLT-PL 04 12-S/07640178987470/4-8832, 5-4148, 5-8438, 6-0721, 6-3209; JLT-PL 04 14-S/07640178987487/4-3861, 4-8833, 5-4149; JLT-PL 04 16-S/07640178987494/5-4150; JLT-PL 04 18-S/07640178987500/5-4151

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1426-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1426-2023Class IIOngoingVoluntary: Firm initiated