Z-1437-2023Class II

Sterile surgical convenience kits: 1. regard Item Number: 800756005, NU00961E - OH Laminectomy Cust Pack; 2...

ROi CPS LLC / initiated 2023-03-22 / Open, Classified / root cause: manufacturing process

ROi CPS LLC began a recall of Sterile surgical convenience kits: 1. regard Item Number: 800756005, NU00961E - OH Laminectomy Cust Pack; 2. regard Item Number: 880411008, NU01018H - Lumbar... on . FDA classified it as Class II and gives the reason as "The recalling firm was notified by their supplier that the light handle covers may separate from the light handle and fall off during use. The light handle covers are a kit component". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1437-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
ROi CPS LLC
Product
Sterile surgical convenience kits: 1. regard Item Number: 800756005, NU00961E - OH Laminectomy Cust Pack; 2. regard Item Number: 880411008, NU01018H - Lumbar Universal ; 3. regard Item Number: 880396007, NU01001G - Craniotomy ; 4. regard Item: 880387005, NU00991E - Ant/Cervical Spine ; and 5. regard Item Number: 800716005, NU00921E - ORMC Major Crani Pack.
Product code
OJG Neurological Tray
Reason for recall
The recalling firm was notified by their supplier that the light handle covers may separate from the light handle and fall off during use. The light handle covers are a kit component.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The recalling firm issued letters dated 3/22/2023 via email the same day. The letter explained the issue, identified the affected product, and listed the actions to be taken by the customer. The actions included checking their inventory for the affected product, segregate and quarantine all affected kits in their inventory, and use the attached label template to print warning labels to be affixed to all affected kits in their inventory. The label should be applied to each affected kit and outer product case in a prominent location to end users of the kits. The location chosen should not cover critical product information found on the existing label. The label instructs users to remove and discard the STERIS Light Handle Cover(s) when the kit is opened for use and to pull a sterile replacement from stock. A reply form is included with the letter for completion to indicate the consignee's actions in response to the notice. The response form is to be returned to the recalling firm via email. If the direct account consignee further distributed or forwarded the affected product, they are to immediately notify those customers. A copy of the ROi CPS recall letter, the letter from STERIS, and the attached label template may be included.
Quantity in commerce
456 kits
Distribution
US Nationwide distribution in the states of FL and MO. There was no foreign/military/government distribution.
Product codes and lots
1. Lot #95370, exp. 10/29/2024, UDI (01)10194717111023(17)241029(10)95370; 2. Lot #95477, exp. 6/30/2024, UDI (01)10194717112242(17)240630(10)95477; 3. Lot #95506, exp. 5/31/2024, UDI (01)10194717114130(17)240531(10)95506; 4. Lot #95632, exp. 5/30/2024, UDI (01)10194717112709(17)240530(10)95632; and 5. Lot #95646, exp. 11/15/2024, UDI (01)10194717110880(17)241115(10)95646, and Lot #95309, exp. 10/29/2024, UDI (01)10194717110880(17)241029(10)95309.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1437-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1431-2023Class IIOngoingVoluntary: Firm initiated
Z-1432-2023Class IIOngoingVoluntary: Firm initiated
Z-1433-2023Class IIOngoingVoluntary: Firm initiated
Z-1434-2023Class IIOngoingVoluntary: Firm initiated
Z-1435-2023Class IIOngoingVoluntary: Firm initiated
Z-1436-2023Class IIOngoingVoluntary: Firm initiated
Z-1437-2023Class IIOngoingVoluntary: Firm initiated
Z-1438-2023Class IIOngoingVoluntary: Firm initiated