ASM Replacement Battery, Component Number SUB0000864
ICU Medical Inc began a recall of ASM Replacement Battery, Component Number SUB0000864 on . FDA classified it as Class I and gives the reason as "Due to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as supplier.
Recall record
- Recall number
- Z-1567-2023
- Classification
- Class I
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Abbott Medical
- Product
- ASM Replacement Battery, Component Number SUB0000864
- Product code
- FRN Pump, Infusion
- Reason for recall
- Due to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the battery lifecycle than expected.
- Root cause tag
- supplier
- FDA root cause
- Under Investigation by firm
- Action
- ICU medical issued an URGENT: MEDICAL DEVICE CORRECTION notice on 03/22/2023 to its direct and indirect consignees by mail. The notice explained the issue and the hazard and requested the following: Clinical Users: Whenever possible, keep the pump plugged into AC power. Before disconnecting the pump from AC power, please ensure that the battery is fully charged. Closely monitor the Battery Status Indicator while the pump is disconnected from AC power. Additionally, have a backup pump available when infusing critical medications. Biomedical Engineering: You may replace affected batteries with a new CSB battery until corrected batteries are available. Please do not use a replacement battery with corroded battery terminals. 1. Identify all affected batteries in your possession and ensure all users or potential users of these pumps are immediately made aware of this notification and proposed mitigations. 2. Complete and return the attached Response Form to [email protected] within ten days of receipt to acknowledge your understanding of this notification. 3. DISTRIBUTORS: If you have distributed potentially affected products to your customers, please immediately forward this notice to them. Request that they complete the response form and return it to [email protected].
- Quantity in commerce
- 9984 units
- Distribution
- worldwide
- Product codes and lots
- Lot Numbers: 6036488, 6827738, 6827739, 6827740, 8408750, 8408755, 8408757
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1565-2023 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-1566-2023 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-1567-2023 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-1568-2023 | Class I | Ongoing | Voluntary: Firm initiated |