Z-1466-2023Class II
stryker ORTHOLOC 3Di Fusion Plate MTP Right, REF 58A220RT, bone stabilization and fixation
Wright Medical Technology Inc / initiated 2023-04-10 / Open, Classified / root cause: manufacturing process
Wright Medical Technology, Inc. began a recall of stryker ORTHOLOC 3Di Fusion Plate MTP Right, REF 58A220RT, bone stabilization and fixation on . FDA classified it as Class II and gives the reason as "Two lots of Ortholoc 3Di Fusion plates contain incorrect dorsiflexion descriptions laser marked on the parts". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1466-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Wright Medical Technology Inc
- Product
- stryker ORTHOLOC 3Di Fusion Plate MTP Right, REF 58A220RT, bone stabilization and fixation
- Product code
- HRS Plate, Fixation, Bone
- Reason for recall
- Two lots of Ortholoc 3Di Fusion plates contain incorrect dorsiflexion descriptions laser marked on the parts.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- The firm issued an URGENT: MEDICAL DEVICE RECALL notice to its sales representatives and distributors on 04/10/2023 by email and followed with letters disseminated to the remaining consignees via tracked FedEx mailing on Wednesday, April 12. The notice explained the issue and the risk and requested the following actions be taken: Identify all affected products and isolate/quarantine to prevent accidental use. All parties that further distributed the product were directed to notify anyone to the recipients of the affected products. The firm is seeking return of the products.
- Quantity in commerce
- 41 units
- Distribution
- US Nationwide distribution.
- Product codes and lots
- UDI/DI 00889797101011, Lot Number 1736809
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1466-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1467-2023 | Class II | Ongoing | Voluntary: Firm initiated |