Aesculap Caiman Articulating D5/360MM - An electrosurgical cutting and coagulation device for...
Aesculap Implant Systems LLC began a recall of Aesculap Caiman Articulating D5/360MM - An electrosurgical cutting and coagulation device for cutting/coagulation/ablation during endoscopic (including... on . FDA classified it as Class II and gives the reason as "Potential for the spring to become loose and fall out of the device, if the spring falls out during a surgical procedure, it may enter the patient's body". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1569-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Aesculap Inc
- Product
- Aesculap Caiman Articulating D5/360MM - An electrosurgical cutting and coagulation device for cutting/coagulation/ablation during endoscopic (including laparoscopic, arthroscopic) surgery, and may in addition be intended for open surgery. Product Code: PL771SU
- Product code
- GEI Electrosurgical, Cutting & Coagulation & Accessories
- Reason for recall
- Potential for the spring to become loose and fall out of the device, if the spring falls out during a surgical procedure, it may enter the patient's body.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Aesculap Inc. issued via Urgent Medical Device Recall Notification Letter via U.S Postal Service, Certified Mail Return Receipt Requested/ USPS Priority Mail on 3/21/23. Letter states reason for recall, health risk and action to take: Review the Device Recall Notification in its entirety and ensure that all users in your organization of the above-mentioned product, and other concerned persons, are informed about this voluntary product recall. If you are a distributor, please forward this recall notification to your customers. The recall is to be extended to the end user/customer level. 2. Determine your current inventory of the affected lots within your facility, cease use and quarantine product subject to recall. Do not destroy any affected product. 3. Utilizing the attached Product Recall Acknowledgement Form, record the total number of individual units in your possession. If you have no inventory remaining, please enter zero (0) on the form. 4. Return the completed Product Recall Acknowledgment Form form to the B. Braun Medical Inc. Quality Assurance department by faxing the form to (610) 849-1197 or e-mail to [email protected] within two (2) weeks of receipt, even if the total inventory in your possession is zero (0). 5. Once we receive your Acknowledgement Form, an Aesculap Customer Support representative will contact you to arrange for credit and to have your current inventory returned. Should you identify a problem or safety concern with the product, please report the event promptly by contacting our Postmarket Surveillance Department at 1-833-425-1464.
- Quantity in commerce
- 6 pieces
- Distribution
- US Nationwide distribution in the state of OH.
- Product codes and lots
- UDI-DI: 04046964837340 Lot Number: 52785146
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1569-2023 | Class II | Ongoing | Voluntary: Firm initiated |