Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1616-2023Class III

To be used with ultrasound endoscopes for ultrasound guided fine needle aspiration (FNA) of submucosal and...

OLYMPUS Corporation of the Americas / initiated 2023-03-31 / Open, Classified / root cause: manufacturing process

Olympus Corporation of the Americas began a recall of To be used with ultrasound endoscopes for ultrasound guided fine needle aspiration (FNA) of submucosal and extramural lesions of the tracheobronchial tree and... on . FDA classified it as Class III and gives the reason as "Product non-conformance pertaining to the angle of the needle does not meet manufacturing specifications and remained unbent, resulting to slight resistance experienced during needle advancement and retraction". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1616-2023
Classification
Class III
Status
Open, Classified
Initiated
Posted
Recalling firm
OLYMPUS Corporation of the Americas
Product
To be used with ultrasound endoscopes for ultrasound guided fine needle aspiration (FNA) of submucosal and extramural lesions of the tracheobronchial tree and the gastrointestinal tract Models: NA-U401SX-4022, NA-U401SX-4022-A
Product code
FCG Biopsy Needle
Reason for recall
Product non-conformance pertaining to the angle of the needle does not meet manufacturing specifications and remained unbent, resulting to slight resistance experienced during needle advancement and retraction.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
Olympus issued Urgent Medical Medical Device Recall letter on 3/31/23 to Endoscopy Department, Pulmonology Department; Risk Management. . Letter states reason for recall, health risk and action to take: 1. Immediately locate any affected products listed in this communication and quarantine these products. 2. Contact our Olympus Customer Service Team at 1-800-848-9024, option 2. Olympus will issue a Return Material Authorization to return any affected product at no charge to you. Olympus will issue a credit to your facility upon return of affected product. 3. Olympus requests that you acknowledge receipt of this letter through the Olympus recall portal. " Go to https://olympusamerica.com/recall and enter the recall number 0422 4. If you have distributed these devices outside your facility, please notify your customers of this matter immediately. Your notification to your customers may be enhanced by including a copy of this letter. Olympus requests that you report complaints, including patient injuries and adverse events, to Olympus. Please report complaints to our Technical Assistance Center (TAC) at 1-800-848-9024, option 1, to report complaints
Quantity in commerce
1297 (Boxes; 5 per Box)
Distribution
Natiowide Foreign: Canada, Germany
Product codes and lots
UDI: 04953170388286 Lot numbers: KR248740 KR248746 KR252863 KR253666 KR253681 KR260854 KR260865 KR260873 KR260875 KR260876 KR260877 KR260896 KR260917 KR260937

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1616-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1615-2023Class IIIOngoingVoluntary: Firm initiated
Z-1616-2023Class IIIOngoingVoluntary: Firm initiated