Distractor Rack, REF 5598007, which may be packaged in Modular Distractor Set, Loaner Kit Number SPS03139...
Medtronic Sofamor Danek USA Inc began a recall of Distractor Rack, REF 5598007, which may be packaged in Modular Distractor Set, Loaner Kit Number SPS03139; for use in spinal surgery on . FDA classified it as Class II and gives the reason as "Two lots of the 5598007 Distractor Rack which were not manufactured to specification. The crest width of the gear is too wide and may cause difficulty or the inability to operate the Distractor". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1574-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Medtronic Sofamor Danek USA Inc
- Product
- Distractor Rack, REF 5598007, which may be packaged in Modular Distractor Set, Loaner Kit Number SPS03139; for use in spinal surgery
- Product code
- LXH Orthopedic Manual Surgical Instrument
- Reason for recall
- Two lots of the 5598007 Distractor Rack which were not manufactured to specification. The crest width of the gear is too wide and may cause difficulty or the inability to operate the Distractor.
- Root cause tag
- manufacturing process
- FDA root cause
- Under Investigation by firm
- Action
- Medtronic issued an URGENT; Medical Device Removal notice on 11/19/2021. The notice explained the issue and the risk and requested the removal of the affected product from use pending return. It also directed the consignee to transfer the notice to other organizations which could be affected.
- Quantity in commerce
- 4 units
- Distribution
- US: IN, CT, NE
- Product codes and lots
- UDI/DI (GTIN) 00763000401535, Lot Numbers EM21D006, EM21D007; Loaner Kit Lot numbers: 0072, 0073, 0077, 0098, 0103
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1574-2023 | Class II | Ongoing | Voluntary: Firm initiated |