Z-1647-2023Class II

Philips Azurion System with software versions: R2.2.O, R2.2.1, R2.2.3, R2.2.5 and R2.2.6. Models/Model...

Philips Medical Systems Nederland B.V. / initiated 2023-03-21 / Open, Classified

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. began a recall of Philips Azurion System with software versions: R2.2.O, R2.2.1, R2.2.3, R2.2.5 and R2.2.6. Models/Model Number: (1) Azurion 3M12: 722063, 722221; (2) Azurion... on . FDA classified it as Class II and gives the reason as "Potential Loss of X-ray-When the full disk capacity is reached, X-Ray functionality will cease to be available there will be a sudden interruption of the procedure". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-1647-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Philips Medical Systems Nederland B.V.
Product
Philips Azurion System with software versions: R2.2.O, R2.2.1, R2.2.3, R2.2.5 and R2.2.6. Models/Model Number: (1) Azurion 3M12: 722063, 722221; (2) Azurion 3M15: 722064, 722222; (3) Azurion SM12: 722227; (4) Azurion SM20: 722228; (5) Azurion 7B12/12: 722067, 722225; (6) Azurion 7820/15: 722068, 722226; (7) Azurion 7M12: 722078, 722223; (8) Azurion 7M20 : 722079, 722224
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
Potential Loss of X-ray-When the full disk capacity is reached, X-Ray functionality will cease to be available there will be a sudden interruption of the procedure
FDA root cause
Under Investigation by firm
Action
Philips issued Urgent Medical Device Letter to US consignees via e-mail on March 21, 2023 and certified email on March 29, 2023. Letters outside of the U.S.A. will be distributed through the local Philips Organizations. Letter states reason for recall, health risk and action to take: Keep this Medical Device Correction with the documentation of the system until Philips corrects your system. Circulate this notice to all users of the system so that they are aware of the issue. Return the attached reply form to Philips to confirm that the users of the system have reviewed and understood this Medical Device Correction. US: Please return this signed form via email to: [email protected] or alternatively via regular mail to Philips at 222 Jacobs Street, Cambridge, MA 02141 for the attention of Mr. Roland Telson / 3rd Floor Philips is working on a software release that will correct this issue (reference FCO72200528). In the interim, until this software is available and installed in your affected system(s), Philips will be removing the Log Trace Files from the affected systems to free up disk capacity (reference FCO72200529). Philips will be prioritizing these activities based on the time the affected software release has been installed in the Philips Azurion system. You will be contacted by your local Philips representative to schedule these activities. If you need any further information or support concerning this issue, contact your local Philips representative.
Quantity in commerce
USA: 1063 R.O.W. 2136
Distribution
Nationwide Foreign: Algeria Argentina Armenia Australia Austria Bahrain Bangladesh Belarus Belgium Bolivia Brazil Bulgaria Cambodia Chile China Colombia Costa Rica C¿te D'Ivoire Croatia Cyprus Czech Republic Denmark Dominican Rep Ecuador Egypt Estonia Finland France French Polynesia Germany Greece Hong Kong Hungary India Indonesia Iran Iraq Ireland Israel Italy Japan Jordan Kazakhstan Kenya Korea, Republic of Kosovo Kuwait Latvia Lebanon Lithuania Macao Macedonia Malaysia Mexico Morocco Myanmar Nepal Netherlands New Zealand Nicaragua Nigeria Norway Oman Pakistan Palestine Palestine, State of Panama Peru Philippines Poland Portugal Puerto Rico Qatar Romania Russian Federation Saudi Arabia Senegal Serbia Singapore Slovakia Slovenia South Africa South Korea Spain Sri Lanka Sweden Switzerland Taiwan Thailand Tunisia Turkey Turkmenistan Ukraine United Kingdom Utd.Arab.Emir. Uzbekistan Vietnam Yemen
Product codes and lots
UDI-DI: System product name Model number UDI (1)Azurion 3M12: 722063 (01)00884838085275(21) 722221 (01)00884838099203(21); (2) Azurion 3M15: 722064 (01)00884838085282(21) 722222 (01)00884838099210(21); (3) Azurion SM12: 722227 (01)00884838099227(21); (4) Azurion SM20: 722228 (01)00884838099234(21); (5) Azurion 7B12/12: 722067 (01)00884838085350(21) 722225 (01)00884838099265(21); (6) Azurion 7820/15: 722068 (01)00884838085367(21) 722226 (01)00884838099272(21); (7) Azurion 7M12: 722078 (01)00884838085251(21) 722223 (01)00884838099241(21); (8) Azurion 7M20: 722079 (01)00884838085268(21) 722224 (01)00884838099258(21), Software: R2.2.O, R2.2.1, R2.2.3, R2.2.5 and R2.2.6

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1647-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1647-2023Class IIOngoingVoluntary: Firm initiated