Z-2261-2023Class II
Incisive CT Plus-Computed Tomography X-Ray System Model: 728149
Philips North America LLC / initiated 2023-04-25 / Open, Classified
Philips North America began a recall of Incisive CT Plus-Computed Tomography X-Ray System Model: 728149 on . FDA classified it as Class II and gives the reason as "Metal mounting box on the rotating scanner on rotor (heat exchanger box) located within the Incisive CT system may become expelled and make contact with other components and cause injury to user or if detached a loud...". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-2261-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Philips North America LLC
- Product
- Incisive CT Plus-Computed Tomography X-Ray System Model: 728149
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- Metal mounting box on the rotating scanner on rotor (heat exchanger box) located within the Incisive CT system may become expelled and make contact with other components and cause injury to user or if detached a loud noise will be emitted and the system will shut down
- FDA root cause
- Under Investigation by firm
- Action
- Phiips issued Urgent Medical Device Correction Letters mailed to US consignees via Certified United States Postal Service (USPS) with delivery confirmation on 4/25/23 International global markets were notified of the issue and instructed to distribute the FSN to affected consignees in accordance with local requirements. Letter states reason for recall, health risk and action to take: Please continue to use your system in accordance with its intended use. If you need to remain in the room for the duration of a scan, Philips recommends using protective eyewear. " Circulate this Urgent Medical Device Correction Letter to all users of this device so that they are aware of the issue. Please retain this letter with your system(s) until a solution is installed on your system; ensure the letter is in a place likely to be seen/viewed. " Place this URGENT Medical Device Correction Letter with your system documentation. " Please complete and return the attached response form to Philips promptly and no later than 30 days from receipt. Completing this form confirms receipt of the Urgent Medical Device Correction Letter, understanding of the issue, and required actions to be taken. Philips will contact you to schedule a time for a Field Service Engineer (FSE) to visit your site and install a solution to address the issue (reference FCO72800798). If you need additional information or support concerning this issue, please contact the Customer Care Solutions Center (1-800-722-9377).
- Quantity in commerce
- 6 units OUS
- Distribution
- Nationwide Foreign: China Algeria Argentina Australia Austria Azerbaijan Bahrain Bangladesh Belarus Belgium Bolivia Brazil Bulgaria Canada Chile China Colombia Czech Republic Ecuador Egypt France Germany Greece India Indonesia Iran Iraq Israel Italy Japan Jordan Korea, Republic of Kosovo Latvia Macedonia Mexico Moldova Netherlands Nigeria Pakistan Panama Philippines Poland Romania Russian Federation Saint Lucia Saint Pierre and Miquelon Saudi Arabia Serbia Slovakia Spain Sweden Taiwan Thailand Tunisia Turkey Turkmenistan United Kingdom Uzbekistan Viet Nam
- Product codes and lots
- UDI-DI: (01)00884838103474(21) Serial Numbers: 530408 530437 530474 530320 530332 530430
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2258-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2259-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2260-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2261-2023 | Class II | Ongoing | Voluntary: Firm initiated |