Teleflex HORIZON Microclip Titanium Ligating Clips w/tape, REF 005200; for use in general surgical procedures...
TELEFLEX LLC began a recall of Teleflex HORIZON Microclip Titanium Ligating Clips w/tape, REF 005200; for use in general surgical procedures that require vessel or tissue ligation on . FDA classified it as Class II and gives the reason as "Teleflex LLC is initiating this voluntary field corrective action because the clip cartridges for the affected product is missing a key feature that restricts the application of the clip to the applier. With this step...". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-1663-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Teleflex Medical
- Product
- Teleflex HORIZON Microclip Titanium Ligating Clips w/tape, REF 005200; for use in general surgical procedures that require vessel or tissue ligation
- Product code
- FZP Clip, Implantable
- Reason for recall
- Teleflex LLC is initiating this voluntary field corrective action because the clip cartridges for the affected product is missing a key feature that restricts the application of the clip to the applier. With this step feature missing on the cartridge, the clip is mounted in a more variable position, allowing the clips to load farther back in the clip applier jaw grooves. This may potentially result in inappropriate or inadequate clip application to vascular structures.
- FDA root cause
- Under Investigation by firm
- Action
- Teleflex issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 04/28/2023 by FedEx 2nd day mail. The letter explained the issue, potential risk, and requested the following actions: Medical facilities: Cease use and distribution, and isolate the affected product. Circulate the notice to all affected departments. Distributors: Cease use and distribution, and isolate the affected product. Provide the notice to your customers. Teleflex is seeking return of the affected units.
- Quantity in commerce
- 146,606 units
- Distribution
- worldwide
- Product codes and lots
- UDI/DI 24026704696394, Lot Numbers: 73A2200857, 73A2300018, 73A2300176, 73A2300326, 73A2300520, 73A2300778, 73D2200216, 73D2200559, 73E2200634, 73E2200874, 73E2201099, 73F2200193, 73F2200200, 73F2200482, 73F2200711, 73G2200168, 73G2200345, 73G2200560, 73G2200754, 73H2200115, 73H2200294, 73H2200433, 73H2200453, 73J2200747, 73J2200763, 73K2200014, 73K2200095, 73K2200101, 73K2200299, 73K2200566, 73K2200713, 73L2200064, 73L2200066
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1663-2023 | Class II | Ongoing | Voluntary: Firm initiated |