Z-1991-2023Class II

Metal Tray, 178 x 135 x 37 mm, REF: 11580A; Metal Tray, 275 x 178 x 35 mm, REF: 11580B; Metal Tray, 373 x 178...

Karl Storz Endoscopy-America, Inc. / initiated 2023-04-06 / Open, Classified / root cause: labeling and UDI

Karl Storz Endoscopy began a recall of Metal Tray, 178 x 135 x 37 mm, REF: 11580A; Metal Tray, 275 x 178 x 35 mm, REF: 11580B; Metal Tray, 373 x 178 x 35 mm, REF: 11580C; Metal Tray, 275 x 175 x 37... on . FDA classified it as Class II and gives the reason as "Metal sterilization trays used to encase and protect medical devices during sterilization have a new 510(k) with instructions for use that is for the first time specific to the tray models, that also includes...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1991-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Karl Storz Endoscopy-America, Inc.
Product
Metal Tray, 178 x 135 x 37 mm, REF: 11580A; Metal Tray, 275 x 178 x 35 mm, REF: 11580B; Metal Tray, 373 x 178 x 35 mm, REF: 11580C; Metal Tray, 275 x 175 x 37 mm, REF: 11580D; Wire Tray f. Cleaning, 290x60x52mm, REF: 39501A1; Wire Tray f. Cleaning, 352 x 125 x 54 mm, REF: 39501A2; Wire Tray f. Cleaning, 487 x 125 x 54 mm, REF: 39501B2; Wire Tray for ENDOCAMELEON, REF: 39501BEC; Tray, with lid, 670x80x52mm, REF: 39501C; Wire Tray for ENDOCAMELEON, long version, REF: 39501CEC; Wire Tray Retromolar Intub. Endoscope, REF: 39501F; Wire basket for c-mac videolaryngoscope, REF: 39501LC2; Tray, URS, 644 x 150 x 80 mm, REF: 39501X; Wire Tray, URS, REF: 39501XK; Wire tray PECD, REF: 39501XP; Tray, paranasal sinus shaver system, REF: 39550A
Product code
KCT Sterilization Wrap Containers, Trays, Cassettes & Other Accessories
Reason for recall
Metal sterilization trays used to encase and protect medical devices during sterilization have a new 510(k) with instructions for use that is for the first time specific to the tray models, that also includes information regarding the sterilization modality.
Root cause tag
labeling and UDI
FDA root cause
No Marketing Application
Action
On 4/6/23, correction notices were distributed to customers who were informed the following: Affected metal trays are intended only for use to encase and protect specific KARL STORZ reusable medical devices for sterilization in pre-vacuum steam sterilization cycles (132C for 4 minutes with a 20-minute dry time). When used in conjunction with an FDA-cleared sterilization wrap, sterility of the enclosed medical device is maintained until used. The system has been validated with devices with working channels greater-than-or-equal-to 0.2 mm and a maximum length of 240 mm. 1) Ensure that all relevant personnel in your organization review this letter, including the new Instructions for Use for the metal trays: following: https://spwebspace.karlstorz.com/sites/MetalTrays/SitePages/Home.aspx 2) If you have previously sold or transferred any of the affected metal trays to a third party, please forward this notice to all such third parties. 3) Complete and return the Acknowledgment Form to [email protected] Customers correction-related questions can contact [email protected] or 888-943-5128. Customers with device-related questions can call 800-421-0837
Quantity in commerce
4,933
Distribution
Worldwide - US Nationwide distribution including in the states of IN, MS, MN, AZ, FL, IL, CA, AK, NY, VA, TX, PA, LA, ND, AS, MD, NH, SC, AR, GA, NJ, NC, WA, WI, SD, TN, KY, AL, OK, MO, MA, MI, CT, UT, WY, CO, OH, DC, DE, NM, MT, ME, OR, PA, WV, IA, NE, HI, KS, ID, NV, RI, VT, GU and the country of CA.
Product codes and lots
REF/UDI-DI: 11580A/ 04048551051801; 11580B/ 04048551051818; 11580C/ 04048551051825; 11580D/ 04048551051832; 39501A1/ 04048551157619; 39501A2/ 04048551157626; 39501B2/ 04048551157657; 39501BEC/ 04048551241400; 39501C/ 04048551157671; 39501CEC/ 04048551254158; 39501F/ 04048551157695; 39501LC2/ 04048551273432; 39501X/ 04048551157756; 39501XK/ 04048551157763; 39501XP/ 04048551157770; 39550A/ 04048551157879. All lots.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1991-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1991-2023Class IIOngoingVoluntary: Firm initiated