Z-1665-2023Class II

Elekta Medical Linear Accelerator (ELMA) included in the below platforms: a. Elekta Infinity; b. Versa HD; c...

Elekta Limited / initiated 2023-05-05 / Open, Classified

Elekta, Inc. began a recall of Elekta Medical Linear Accelerator (ELMA) included in the below platforms: a. Elekta Infinity; b. Versa HD; c. Elekta Synergy on . FDA classified it as Class II and gives the reason as "A component within a limited number of Elekta Linacs has been realized to be non-compliant to IEC 60601". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-1665-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Elekta Limited
Product
Elekta Medical Linear Accelerator (ELMA) included in the below platforms: a. Elekta Infinity; b. Versa HD; c. Elekta Synergy
Product code
IYE Accelerator, Linear, Medical
Reason for recall
A component within a limited number of Elekta Linacs has been realized to be non-compliant to IEC 60601
FDA root cause
Under Investigation by firm
Action
Elekta notified customers, via email, on 05/05/2023. Customers were instructed that Elekta will contact each impacted customer to perform an onsite inspection of the power supply unit and replace as needed. Customers were asked to complete and return the Acknowledgement Form.
Quantity in commerce
46 units
Distribution
Worldwide distribution. US distribution to AR, IN, LA, OK, TN, VA, WA. International distribution to Belgium, Canada, Chile, China, France, Germany, India, Iraq, Italy, Jordan, Libyan Arab Jamahiriya, Mauritius, Mexico, Morocco, Poland, Russian Federation, Spain.
Product codes and lots
Varex Part Number 95510354H-01 (Varex serial numbers 354-01H_A0010004PC to 354-01H_A0010269PC). a. Elekta Infinity: UDI-DI Number: 05060191071543, 05060191071512 (Serial Numbers: 156073, 152223, 153681); b. Versa HD: UDI-DI Number: 05060191071574, 05060191071529 (Serial Numbers: 156941, 156989, 156724, 156998); c. Elekta Synergy: UDI-DI Number: 05060191071550, 05060191071505 (Serial Number: 152483) Additional serial numbers distributed OUS: 109331, 109333, 109335, 109336, 109339, 109341, 109342, 151879, 152194, 153038, 153299, 153410, 153970, 156841, 156849, 156966, 156968, 156969, 156970, 156971, 156973, 156977, 156978, 156979, 156980, 156981, 156982, 156983, 156984, 156985, 156986, 156987, 156988, 156990, 156992, 156995, 156996, 156997.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1665-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1665-2023Class IIOngoingVoluntary: Firm initiated