Z-1696-2023Class III
Alinity i Anti-TPO Reagent Kit, List Number 09P3521
Abbott Ireland / initiated 2023-04-24 / Open, Classified / root cause: manufacturing process
Abbott Ireland Limited began a recall of Alinity i Anti-TPO Reagent Kit, List Number 09P3521 on . FDA classified it as Class III and gives the reason as "Due to a manufacturing fill volume error, Alinity i Anti-TPO Reagent contains a marginally lower fill volume than detailed in the Kit Contents section of the Instructions for Use (IFU)". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1696-2023
- Classification
- Class III
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Abbott Ireland
- Product
- Alinity i Anti-TPO Reagent Kit, List Number 09P3521
- Product code
- JZO System, Test, Thyroid Autoantibody
- Reason for recall
- Due to a manufacturing fill volume error, Alinity i Anti-TPO Reagent contains a marginally lower fill volume than detailed in the Kit Contents section of the Instructions for Use (IFU).
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- On April 24, 2023, Abbott issued a Product Correction letter, via FedEx priority overnight express, for this issue. Customers were instructed to provide a copy of the Product Correction letter to the laboratory manager, supervisor or health professional responsible for the impacted product and to complete and return the Customer Reply form.
- Quantity in commerce
- 110 kits
- Distribution
- US Nationwide distribution in the states of CA, CO, FL, GA, IL, KY, LA, MD, MI, MN, NJ, NY, OH, RI, SC, SD, TN, TX, WA. No international distribution of affected lot.
- Product codes and lots
- UDI Number: (01)00380740153892(17)240720(10)46009FN00; List Number: 09P3521; Lot Number: 46009FN00; Expiration Date: 07/20/2024
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1696-2023 | Class III | Ongoing | Voluntary: Firm initiated |