Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1696-2023Class III

Alinity i Anti-TPO Reagent Kit, List Number 09P3521

Abbott Ireland / initiated 2023-04-24 / Open, Classified / root cause: manufacturing process

Abbott Ireland Limited began a recall of Alinity i Anti-TPO Reagent Kit, List Number 09P3521 on . FDA classified it as Class III and gives the reason as "Due to a manufacturing fill volume error, Alinity i Anti-TPO Reagent contains a marginally lower fill volume than detailed in the Kit Contents section of the Instructions for Use (IFU)". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1696-2023
Classification
Class III
Status
Open, Classified
Initiated
Posted
Recalling firm
Abbott Ireland
Product
Alinity i Anti-TPO Reagent Kit, List Number 09P3521
Product code
JZO System, Test, Thyroid Autoantibody
Reason for recall
Due to a manufacturing fill volume error, Alinity i Anti-TPO Reagent contains a marginally lower fill volume than detailed in the Kit Contents section of the Instructions for Use (IFU).
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
On April 24, 2023, Abbott issued a Product Correction letter, via FedEx priority overnight express, for this issue. Customers were instructed to provide a copy of the Product Correction letter to the laboratory manager, supervisor or health professional responsible for the impacted product and to complete and return the Customer Reply form.
Quantity in commerce
110 kits
Distribution
US Nationwide distribution in the states of CA, CO, FL, GA, IL, KY, LA, MD, MI, MN, NJ, NY, OH, RI, SC, SD, TN, TX, WA. No international distribution of affected lot.
Product codes and lots
UDI Number: (01)00380740153892(17)240720(10)46009FN00; List Number: 09P3521; Lot Number: 46009FN00; Expiration Date: 07/20/2024

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1696-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1696-2023Class IIIOngoingVoluntary: Firm initiated