Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1633-2023Class I

Walnut Wearable Smart Thermometer, intended for continuous chest temperature monitoring of children ages 0-6...

BearCare, Inc. / initiated 2023-05-10 / Terminated / root cause: design

BearCare, Inc. began a recall of Walnut Wearable Smart Thermometer, intended for continuous chest temperature monitoring of children ages 0-6 years in non-emergency medical situations, Model... on . FDA classified it as Class I and gives the reason as "There have been reports of skin burns to children wearing the device". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1633-2023
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
BearCare, Inc.
Product
Walnut Wearable Smart Thermometer, intended for continuous chest temperature monitoring of children ages 0-6 years in non-emergency medical situations, Model Number WT20
Product code
FQZ Thermometer, Clinical Color Change
Reason for recall
There have been reports of skin burns to children wearing the device.
Root cause tag
design
FDA root cause
Component design/selection
Action
An URGENT: MEDICAL DEVICE RECALL notification letter dated 5/10/23 was sent to customers. Actions to take: BearCare is asking consumers to immediately discontinue use of all Walnut Thermometers employing a rechargeable battery and return the product to BearCare for a refund. BearCare s records indicate approximately nine hundred seventy-nine (979) units are in the possession of consumers. Any consumer noting burns or irritation on their child s skin should seek immediate medical care. Additionally, BearCare is asking consumers to: (1) confirm receipt of the recall notice via email to [email protected] and (2) return all rechargeable Walnut Thermometers to BearCare for evaluation using a prepaid shipping label. BearCare is asking all retailers to immediately remove the rechargeable Walnut Thermometer from all online marketplaces and return the product to BearCare. BearCare s records indicate approximately eight hundred thirty-nine (839) outstanding units are in the possession of retailers. BearCare is vigorously investigating these complaints to determine the root cause of the alleged failure, including working with professional consultants to evaluate the hardware and packaging of the device. BearCare has voluntarily decided to remove the rechargeable Walnut Thermometer from the market to mitigate any further risk to the public. Return Instructions: BearCare is asking consumers to return the product to the address below to obtain a full refund of the purchase price. Consumers may contact [email protected] for a prepaid shipping label. You can find the recall announcement posted at https://walnutcares.com/.
Quantity in commerce
1818 units
Distribution
US Nationwide distribution.
Product codes and lots
Lot Number 20221115W002

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1633-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1633-2023Class ITerminatedVoluntary: Firm initiated