Smith&Nephew ENGAGE Cementless Partial Knee System, Coated Tibial Tray, Part Numbers: a) REF 1-10012-100...
Smith & Nephew, Inc. began a recall of Smith&Nephew ENGAGE Cementless Partial Knee System, Coated Tibial Tray, Part Numbers: a) REF 1-10012-100, SIZE 1-LEFT MEDIAL; b) REF 1-10012-150, SIZE 1-RIGHT... on . FDA classified it as Class II and gives the reason as "Recent complaint data indicates that the revision rate may be trending higher than corresponding similar devices in global joint replacement registries. The data identifies a potential signal that the performance is an...". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-1672-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Smith & Nephew, Inc.
- Product
- Smith&Nephew ENGAGE Cementless Partial Knee System, Coated Tibial Tray, Part Numbers: a) REF 1-10012-100, SIZE 1-LEFT MEDIAL; b) REF 1-10012-150, SIZE 1-RIGHT MEDIAL; c) REF 1-10012-200, SIZE 2-LEFT MEDIAL; d) REF 1-10012-250, SIZE 2-RIGHT MEDIAL; e) REF 1-10012-300, SIZE 3-LEFT MEDIAL; f) REF 1-10012-350, SIZE 3-RIGHT MEDIAL; g) REF 1-10012-400, SIZE 4-LEFT MEDIAL; h) REF 1-10012-450, SIZE 4-RIGHT MEDIAL; i) REF 1-10012-500, SIZE 5-LEFT MEDIAL; j) REF 1-10012-550, SIZE 5-RIGHT MEDIAL; k) REF 1-10012-600, SIZE 6-LEFT MEDIAL; l) REF 1-10012-650, SIZE 6-RIGHT MEDIAL; Unicondylar knee prothesis
- Product code
- HSX Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer
- Reason for recall
- Recent complaint data indicates that the revision rate may be trending higher than corresponding similar devices in global joint replacement registries. The data identifies a potential signal that the performance is an outlier versus the state of the art with respect to the risk for revision.
- FDA root cause
- Under Investigation by firm
- Action
- Smith & Nephew issued an Urgent Medical Device Recall Notice to its consignees on 05/02/2023 via email and letter. The notice explained the problem and the risk and requested the following: Required Actions: 1. Please inspect your inventory and locate any devices from the listed product and batch numbers enclosed with this Urgent Medical Device Recall Notice, and quarantine them immediately. a. If you are a sales rep, district office, or distributor, you must notify your customers of the field action and ensure that the required actions are complete. b. Include the customer s account number and name in Section A. 2. Please provide a copy of the attached Physician Communication to the surgical staff. The firm is seeking the return of the device.
- Quantity in commerce
- 2146 units
- Distribution
- US Nationwide distribution.
- Product codes and lots
- a) REF 1-10012-100, UDI/DI 00885556873113; b) REF 1-10012-150, UDI/DI 00885556873120; c) REF 1-10012-200, UDI/DI 00885556873137; d) REF 1-10012-250, UDI/DI 00885556873144; e) REF 1-10012-300, UDI/DI 00885556873151; f) REF 1-10012-350, UDI/DI 00885556873168; g) REF 1-10012-400, UDI/DI 00885556873175; h) REF 1-10012-450, UDI/DI 00885556873182; i) REF 1-10012-500, UDI/DI 00885556873199; j) REF 1-10012-550, UDI/DI 00885556873205; k) REF 1-10012-600, UDI/DI 00885556873212; l) REF 1-10012-650, UDI/DI 00885556873229; ALL BATCHES
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1670-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1671-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1672-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1673-2023 | Class II | Ongoing | Voluntary: Firm initiated |