Z-2007-2023Class II

NaviCare Patient Safety V3.9.200 to 3.9.600, used with the Centrella Bed

Baxter Healthcare Corporation / initiated 2023-05-19 / Open, Classified / root cause: design

Baxter Healthcare Corporation began a recall of NaviCare Patient Safety V3.9.200 to 3.9.600, used with the Centrella Bed on . FDA classified it as Class II and gives the reason as "Baxter identified a potential risk where the "safety" monitoring and "bed exit" monitoring features may enter a permanent state of alert suppression. In this state, a caregiver may not be alerted remotely when assigned...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-2007-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Baxter Healthcare Corporation
Product
NaviCare Patient Safety V3.9.200 to 3.9.600, used with the Centrella Bed
Product code
FNL Bed, Ac-Powered Adjustable Hospital
Reason for recall
Baxter identified a potential risk where the "safety" monitoring and "bed exit" monitoring features may enter a permanent state of alert suppression. In this state, a caregiver may not be alerted remotely when assigned patient protocols (bed exit, siderail position, brakes off, height of bed changes, head of bed angle, and/or turn reminders) are turned off.
Root cause tag
design
FDA root cause
Device Design
Action
Baxter issued an URGENT MEDICAL DEVICE CORRECTION NOTICE to its consignees on 05/19/2023 by letter (USPS). The notice explained the issue, hazard, and requested the following: Baxter is asking that you take the following actions: 1. Share this letter with your IT department (or other relevant personnel) to determine if you have the impacted software version and contact Baxter Technical Support at 1-800-445-3720, available twenty-four hours a day, seven days a week, to schedule a software upgrade which addresses the identified issue. 2. You may continue to use your Voalte Patient Safety Software (VPS) (with your Centrella bed) until a software upgrade is completed by implementing either of these options into your clinical bedside workflow: Option 1: When working with a patient in the room, and a bed exit alarm is activated, only silence the bed, and extend the silence timer instead of turning off bed exit. " At the Graphical Room Station (GRS), click Menu " Click Manage Bed " On the Manage Bed screen, click the Safety Alerts ON / OFF slider button and the Delay button will pop up. Extend the silence period for the desired amount of time. Option 2: After working with a patient in the room, and prior to exiting the room, enable Patient Safety at the Graphical Room Station (GRS). " At the Graphical Room Station, click Menu " Click Manage Bed on the Manage Bed screen, toggle the Safety Alerts button to ON . For general questions regarding this communication, contact Baxter Technical Support at 1-800-445-3720, available twenty-four hours a day, seven days a week.
Quantity in commerce
7 software versions
Distribution
US Nationwide distribution in the states of OR, SC, VA, TN, GA, NC, CO, NC, FL.
Product codes and lots
UDI/DI 00887761985193, Software Versions: 3.9: 3.9.200,3.9.201, 3.9.300, 3.9.400, 3.9.401, 3.9.500, 3.9.600

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2007-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2006-2023Class IIOngoingVoluntary: Firm initiated
Z-2007-2023Class IIOngoingVoluntary: Firm initiated