MEDLINE THERAHONEY GEL, 100% LEPTOSPERMUM SCOPARIUM HONEY, a) REF MNK0005, b) REF MNK0015 wound gel
MEDLINE INDUSTRIES, LP - Northfield began a recall of MEDLINE THERAHONEY GEL, 100% LEPTOSPERMUM SCOPARIUM HONEY, a) REF MNK0005, b) REF MNK0015 wound gel on . FDA classified it as Class II and gives the reason as "Medline issued a recall on TheraHoney Wound Gel due to the manufacturer notifying Medline of a sterilization dose audit failure on the wound gel and not augmenting the dose after failure per ISO 11137". As of , the recall status on record is Open, Classified, and MDM tags the root cause as sterility.
Recall record
- Recall number
- Z-1982-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Product
- MEDLINE THERAHONEY GEL, 100% LEPTOSPERMUM SCOPARIUM HONEY, a) REF MNK0005, b) REF MNK0015 wound gel
- Product code
- FRO Dressing, Wound, Drug
- Reason for recall
- Medline issued a recall on TheraHoney Wound Gel due to the manufacturer notifying Medline of a sterilization dose audit failure on the wound gel and not augmenting the dose after failure per ISO 11137.
- Root cause tag
- sterility
- FDA root cause
- Under Investigation by firm
- Action
- Medline issued a recall notice to its consignees on 05/03/2023 by first calls mail and email The notice explained the problem with the device, risk and requested the following: "REQUIRED ACTION: 1.Immediately check your stock for the affected item number and the affected lot numbers listed on the recall portal. Destroy affected product.2.Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. The login for completing the response form is: Website link: https://recalls.medline.com Recall Reference #: R-23-059 Recall Code: RXXXXXXX 3.Upon completion of the form, please destroy affected product. Your account will receive credit once the response form is submitted. 4.If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Have the customers document and destroy any affected product. You should include your customers quantities on your response form."
- Quantity in commerce
- 851,328 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of SINGAPORE, DOMINICAN REPUBLIC, UNITED ARAB EMIRATES, DUBAI, CHINA, PANAMA, CANADA, SAUDI ARABIA, BERMUDA, COLOMBIA, MAURITIUS, KUWAIT.
- Product codes and lots
- a) REF MNK0005: GTIN 30884389161566, Lot Numbers: LMPTH0189, LMPTH0191, LMPTH0193, LMPTH0195, LMPTH0197, LMPTH0199, LMPTH0201, LMPTH0203, LMPTH0205, LMPTH0207, LMPTH0211, LMPTH0213; b) REF MNK0015: GTIN 40884389154558, Lot Numbers: LMPTH0188, LMPTH0190, LMPTH0192, LMPTH0196, LMPTH0198, LMPTH0200, LMPTH0202, LMPTH0204, LMPTH0206, LMPTH0210, LMPTH0210B, LMPTH0212
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1982-2023 | Class II | Ongoing | Voluntary: Firm initiated |