Cardiosave Rescue Intra-Aortic Balloon Pump (IABP), Model numbers 0998-00-0800-75 0998-00-0800-83...

Datascope Corp. / initiated 2023-06-05 / Open, Classified / root cause: design

Datascope Corp. began a recall of Cardiosave Rescue Intra-Aortic Balloon Pump (IABP), Model numbers 0998-00-0800-75 0998-00-0800-83 0998-00-0800-85 on . FDA classified it as Class I and gives the reason as "An unexpected shutdown of the IABP may occur due to failures of tantalum capacitors in the Power Management Board and/or Solenoid Board. This issue may lead to an unexpected interruption of therapy". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-2152-2023
Classification
Class I
Status
Open, Classified
Initiated
Posted
Recalling firm
Datascope Corp.
Product
Cardiosave Rescue Intra-Aortic Balloon Pump (IABP), Model numbers 0998-00-0800-75 0998-00-0800-83 0998-00-0800-85
Product code
DSP System, Balloon, Intra-Aortic And Control
Reason for recall
An unexpected shutdown of the IABP may occur due to failures of tantalum capacitors in the Power Management Board and/or Solenoid Board. This issue may lead to an unexpected interruption of therapy.
Root cause tag
design
FDA root cause
Component design/selection
Action
On June 5, 2023, the firm notified customers of the recall via "URGENT MEDICAL DEVICE CORRECTION" letters. Letters covered two issues: Issue 1 - Failure to charge battery (Covered by RES 92287) Issue 2 - Unexpected shutdown (Covered by this RES event) User Actions: Issue 2: Unexpected Shutdown due to failure of Tantalum Capacitors. Should loss of power occur, therapy will be interrupted and the console will need to be exchanged with another (if available) to resume therapy. If your device remains inoperable, please contact your service representative. It is recommended for users to have an alternate IABP on hand in the event that this issue may occur. The firm is developing a hardware correction to address this issue. If you have any questions, please contact your Datascope/Getinge representative or call Datascope/Getinge Technical Support at 1-888-943-8872, options 4, 2, 1, Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. ET
Quantity in commerce
9174 (4586 US; 4588 OUS) in total
Distribution
Worldwide Distribution- US (Nationwide) and OUS (foreign) countries of: ALBANIA, ALGERIA, ANGOLA, ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, AZERBAIJAN , BAHRAIN, BANGLADESH, BARBADOS, BELARUS, BELGIUM, BOSNIA and HERZEGOVINA, BRAZIL, BRUNEI, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, C¿te d'Ivoire, CROATIA, CZECH REPUBLIC, ECUADOR, EGYPT, EL SALVADOR, ESTONIA, FINLAND, FRANCE, GERMANY, GHANA, GUATEMALA, HONG KONG, HUNGARY, ICELAND, INDIA, INDONESIA, IRAN, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KENYA, KOSOVO, KUWAIT, LATVIA, LEBANON, LIBYA, LITHUANIA, LUXEMBOURG, MACEDONIA, MALAYSIA, MEXICO, MOROCCO, MYANMAR, NETHERLANDS, NEW ZEALAND, NIGERIA, NORWAY, OMAN, PAKISTAN, PALESTINE, PERU, PHILIPPINES, POLAND, PORTUGAL, QATAR, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SRI LANKA, SWEDEN, SWITZERLAND, SYRIA, TAIWAN, TANZANIA, THAILAND, TRINIDAD AND TOBAGO, TUNISIA, TURKEY, TURKMENISTAN, UKRAINE, UNITED ARAB EMIRATES, UNITED KINGDOM, UZBEKISTAN, VIETNAM, YEMEN, ZAMBIA
Product codes and lots
All Unit Serial Numbers Model UDI 0998-00-0800-75 10607567112312 0998-00-0800-83 10607567108407 0998-00-0800-85 10607567113449

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2152-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2151-2023Class IOngoingVoluntary: Firm initiated
Z-2152-2023Class IOngoingVoluntary: Firm initiated