Z-1980-2023Class II
V-Sign Sensor 2, REF VS-A/P/N, a component of Sentec Digital Monitoring System (model SDMS)
Sentec AG / initiated 2023-05-02 / Open, Classified / root cause: design
SenTec AG began a recall of V-Sign Sensor 2, REF VS-A/P/N, a component of Sentec Digital Monitoring System (model SDMS) on . FDA classified it as Class II and gives the reason as "The sensors may experience an out-of-box failure because after recalibration, the sensors stayed in the software mode "production" and were not reset to "released" mode". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1980-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Sentec AG
- Product
- V-Sign Sensor 2, REF VS-A/P/N, a component of Sentec Digital Monitoring System (model SDMS).
- Product code
- LKD Monitor, Carbon-Dioxide, Cutaneous
- Reason for recall
- The sensors may experience an out-of-box failure because after recalibration, the sensors stayed in the software mode "production" and were not reset to "released" mode.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The recalling firm issued a recall letter dated 5/2/2023 via email on 5/2/2023. The letter explained the reason for recall also stating that the sensors pose no risk to patients but have a high chance of not functioning upon being connected to an SDM monitor. The recalling firm informed the consignee they want to replace the defective sensors. An RMA was enclosed to return the sensor and the shipping confirmation of its replacement.
- Quantity in commerce
- 11 in the U.S.
- Distribution
- US Nationwide distribution in the states of CA, FL, IL, MA, MS, MT, NV, and TX.
- Product codes and lots
- Serial numbers 338947, 338952, 338953, 338963, 338967, 338989, 338991, 338993, 338994, 338996, 338997, UDI-DI 07640121880513.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1980-2023 | Class II | Ongoing | Voluntary: Firm initiated |