Z-2014-2023Class I
Olympus EVIS EXERA BRONCHOVIDEOSCOPE BF TYPE, Model Number BF-XT160
OLYMPUS Corporation of the Americas / initiated 2023-06-08 / Open, Classified
Olympus Corporation of the Americas began a recall of Olympus EVIS EXERA BRONCHOVIDEOSCOPE BF TYPE, Model Number BF-XT160 on . FDA classified it as Class I and gives the reason as "There have been complaints of endobronchial combustion during therapeutic laser procedures with the bronchoscope". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-2014-2023
- Classification
- Class I
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- OLYMPUS Corporation of the Americas
- Product
- Olympus EVIS EXERA BRONCHOVIDEOSCOPE BF TYPE, Model Number BF-XT160
- Product code
- EOQ Bronchoscope (Flexible Or Rigid)
- Reason for recall
- There have been complaints of endobronchial combustion during therapeutic laser procedures with the bronchoscope.
- FDA root cause
- Under Investigation by firm
- Action
- An "URGENT MEDICAL DEVICE CORRECTIVE ACTION" notification letter dated 6/8/23 was sent to customers. Actions to be taken by the company: The labelling will be updated to include specificity about laser compatibility, improved instructions regarding patient preparation, and warnings about patient injury and death resulting from incompatible laser use. Actions to be taken by the end user: 1. Inspect your inventory for the referenced devices and identify any device with the model names. Please check all areas of the hospital to determine if any of these devices remain in inventory. 2. Carefully read the content of this Medical Device Correction Action as well as the attached Addendum . The addendum provides compatible laser type. 3. Ensure all personnel are completely knowledgeable and thoroughly aware that Olympus laser compatible bronchoscopes are compatible only with Nd: YAG laser or 810 nm diode lasers. 4. Olympus requests that you acknowledge receipt of this letter through the Olympus recall portal. a. Go to https://olympusamerica.com/recall b. Enter the file (recall) number: 0427 5. If you have further distributed this product, identify your customers, forward them this notification, and appropriately document your notification process. Olympus requests that you report complaints, including any injuries associated with laser procedures with Olympus bronchoscopes, to our Technical Assistance Center (TAC) at 1-800-848-9024 (option 1). Adverse events experienced with the use of this product may also be reported to the Food and Drug Administration MedWatch Adverse Event Reporting program either online, by regular mail or by fax. Olympus regrets any inconvenience caused and fully appreciates your cooperation in this matter. Please do not hesitate to call (647) 999-3203 or email: at [email protected] for any additional information or support concerning this matter.
- Quantity in commerce
- 284 units
- Distribution
- US Nationwide Distribution.
- Product codes and lots
- All serial numbers BF-XT160* EVIS EXERA BRONCHOVIDEOSCOPE OLYMPUS BF TYPE XT160 UDI: 04953170340147
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2011-2023 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-2012-2023 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-2013-2023 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-2014-2023 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-2015-2023 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-2016-2023 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-2017-2023 | Class I | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
