Medtronic ILLUMISITE Platform Console, Model Number ILS-1000-CS
Covidien, LLC began a recall of Medtronic ILLUMISITE Platform Console, Model Number ILS-1000-CS on . FDA classified it as Class II and gives the reason as "Single ILLUMISITE Platform Console was converted from a 220-240VAC model to a 120VAC configuration using a non-supplied power cord and non-supplied fuses in the field by Medtronic Field Service Engineer". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2008-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Medtronic, Inc.
- Product
- Medtronic ILLUMISITE Platform Console, Model Number ILS-1000-CS
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- Single ILLUMISITE Platform Console was converted from a 220-240VAC model to a 120VAC configuration using a non-supplied power cord and non-supplied fuses in the field by Medtronic Field Service Engineer.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Covidien notified the single consignee via letter on 05/18/2023. The consignee was instructed to quarantine and return affected product, forward the notification to those who need to be aware within the organization and to any location which the device has been transferred and complete and return the Customer Confirmation Form.
- Quantity in commerce
- 1 unit
- Distribution
- International distribution in the country of Taiwan.
- Product codes and lots
- UDI-DI: 10884521782211; Serial Number: 518503
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2008-2023 | Class II | Ongoing | Voluntary: Firm initiated |