Z-2008-2023Class II

Medtronic ILLUMISITE Platform Console, Model Number ILS-1000-CS

Medtronic, Inc. / initiated 2023-05-22 / Open, Classified / root cause: manufacturing process

Covidien, LLC began a recall of Medtronic ILLUMISITE Platform Console, Model Number ILS-1000-CS on . FDA classified it as Class II and gives the reason as "Single ILLUMISITE Platform Console was converted from a 220-240VAC model to a 120VAC configuration using a non-supplied power cord and non-supplied fuses in the field by Medtronic Field Service Engineer". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2008-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Medtronic, Inc.
Product
Medtronic ILLUMISITE Platform Console, Model Number ILS-1000-CS
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
Single ILLUMISITE Platform Console was converted from a 220-240VAC model to a 120VAC configuration using a non-supplied power cord and non-supplied fuses in the field by Medtronic Field Service Engineer.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Covidien notified the single consignee via letter on 05/18/2023. The consignee was instructed to quarantine and return affected product, forward the notification to those who need to be aware within the organization and to any location which the device has been transferred and complete and return the Customer Confirmation Form.
Quantity in commerce
1 unit
Distribution
International distribution in the country of Taiwan.
Product codes and lots
UDI-DI: 10884521782211; Serial Number: 518503

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2008-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2008-2023Class IIOngoingVoluntary: Firm initiated