Z-2206-2023Class II

Medline O.R. Scissors, S/B, Sterile, 5.5, Model #DYND04000

MEDLINE INDUSTRIES, LP - Northfield / initiated 2023-05-12 / Open, Classified / root cause: packaging

MEDLINE INDUSTRIES, LP - Northfield began a recall of Medline O.R. Scissors, S/B, Sterile, 5.5, Model #DYND04000 on . FDA classified it as Class II and gives the reason as "The scissors were manufactured without a tip protector resulting in the scissors breaking through the semi-rigid plastic tray compromising sterility". As of , the recall status on record is Open, Classified, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-2206-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Product
Medline O.R. Scissors, S/B, Sterile, 5.5, Model #DYND04000
Product code
HRR Scissors, Orthopedic, Surgical
Reason for recall
The scissors were manufactured without a tip protector resulting in the scissors breaking through the semi-rigid plastic tray compromising sterility.
Root cause tag
packaging
FDA root cause
Under Investigation by firm
Action
The recalling firm issued letters dated 5/12/2023 via first class mail and email on 5/12/2023. The letter explained the reason for recall and provided the required actions. The actions were to immediately check their inventory for the affected item number and lot numbers listed on the recalling firm's recall portal and destroy the products. A website link is provided for completing the response form and for obtaining the specific item and lot numbers affected. Once the response form is submitted, the consignee account will be credited. Distributors are instructed to notify the customers of the recall who have received the affected product and the customer is to document the destruction of their product. The distributor is to include the customer product quantities in their response form.
Quantity in commerce
189,221 scissors
Distribution
US Nationwide distribution. There was government distribution but no military or foreign distribution.
Product codes and lots
Lot numbers and UDI numbers: 23BBI031, 01100801968242291023BBI031; 23BBK353, 01100801968242291023BBK353; 23BBP830, 01100801968242291023BBP830; 23BBQ676, 01100801968242291023BBQ676; 23BBR546, 01100801968242291023BBR546; 23CBH009, 01100801968242291023CBH009; 23CBL930, 01100801968242291023CBL930; 23CBP165, 01100801968242291023CBP165; 23BBS369, 01100801968242291023BBS369; 23BBT718, 01100801968242291023BBT718; 23CBV064, 01100801968242291023CBV064; 23CBW834, 01100801968242291023CBW834; 23CBV222, 01100801968242291023CBV222; 23CBW020, 01100801968242291023CBW020; and 23CBQ190, 01100801968242291023CBQ190.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2206-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2206-2023Class IIOngoingVoluntary: Firm initiated