Z-2138-2023Class II
COULTER DxH Diluent, REF 628017; diluent used for counting and sizing blood cells
Siemens Healthcare Diagnostics Inc. / initiated 2023-05-31 / Open, Classified / root cause: manufacturing process
Beckman Coulter, Inc. began a recall of COULTER DxH Diluent, REF 628017; diluent used for counting and sizing blood cells on . FDA classified it as Class II and gives the reason as "Specific diluent lot numbers showed that conductivity, osmolality, and pH were out of manufacturing specifications". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2138-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- COULTER DxH Diluent, REF 628017; diluent used for counting and sizing blood cells
- Product code
- GIF Diluent, Blood Cell
- Reason for recall
- Specific diluent lot numbers showed that conductivity, osmolality, and pH were out of manufacturing specifications.
- Root cause tag
- manufacturing process
- FDA root cause
- Under Investigation by firm
- Action
- Beckman Coulter issued an URGENT MEDICAL DEVICE RECALL notice on 05/31/2023 by email and mail. The notice explained the issue, its impact on the patient, and requested the following actions be taken: "Inspect the lot number set up on your instrument(s) and your current inventory with the corresponding container number (see Appendix A). Locate the container number on your diluent reagent 2D or Linear barcode label (see Appendix B). Contact Beckman Coulter Customer Technical Support Organization to replace the affected containers in your current inventory. Consult with your laboratory management to determine whether a retrospective review of results is necessary.
- Quantity in commerce
- 335, 448 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Argentina, Bolivia (Plurinational State of), Brazil, Canada, Chile, Colombia, Ecuador, Hong Kong, Indonesia, Kuwait, Mexico, Panama, Peru, Puerto Rico, Taiwan, Trinidad and Tobago, Uruguay, and Viet Nam.
- Product codes and lots
- UDI DI 150995901HEMDILUENT5R; Container Number: 0001-0401: Lot numbers: Lot Numbers Beginning with 3541: 3541500, 3541510, 3541520, 3541550, 3541610, 3541620, 3541670, 3541690, 3541750, 3541770, 3541790, 3541810, 3541830, 3541860, 3541880, 3541900, 3541910, 3541920, 3541930, 3541940, 3541950, 3541990; Lot Numbers Beginning with 3542: 3542010, 3542030, 3542050, 3542070, 3542110, 3542150, 3542170, 3542180, 3542190, 3542230, 3542250, 3542260, 3542270, 3542310, 3542350, 3542370, 3542450, 3542490, 3542510, 3542530, 3542570, 3542590, 3542610, 3542630, 3542650, 3542670, 3542720, 3542740, 3542760, 3542780, 3542800, 3542820, 3542840, 3542850, 3542900, 3542920, 3542940, 3542960, 3542980; Lot Numbers Beginning with 3543: 3543000, 3543020, 3543040, 3543050, 3543080, 3543120, 3543140, 3543160, 3543170, 3543180, 3543200, 3543220, 3543240, 3543280, 3543300, 3543320, 3543340, 3543360, 3543370, 3543390, 3543410, 3543430, 3543460, 3543480, 3543500, 3543540, 3543560, 3543580, 3543610, 3543850, 3543870, 3543890, 3543910, 3543930, 3543950, 3543970, 3543990; Lot Numbers Beginning with 3544: 3544010, 3544030, 3544040, 3544050, 3544070, 3544090, 3544110, 3544130, 3544150, 3544160, 3544170, 3544190, 3544210, 3544230, 3544250
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2138-2023 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.