Z-2138-2023Class II

COULTER DxH Diluent, REF 628017; diluent used for counting and sizing blood cells

Siemens Healthcare Diagnostics Inc. / initiated 2023-05-31 / Open, Classified / root cause: manufacturing process

Beckman Coulter, Inc. began a recall of COULTER DxH Diluent, REF 628017; diluent used for counting and sizing blood cells on . FDA classified it as Class II and gives the reason as "Specific diluent lot numbers showed that conductivity, osmolality, and pH were out of manufacturing specifications". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2138-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
COULTER DxH Diluent, REF 628017; diluent used for counting and sizing blood cells
Product code
GIF Diluent, Blood Cell
Reason for recall
Specific diluent lot numbers showed that conductivity, osmolality, and pH were out of manufacturing specifications.
Root cause tag
manufacturing process
FDA root cause
Under Investigation by firm
Action
Beckman Coulter issued an URGENT MEDICAL DEVICE RECALL notice on 05/31/2023 by email and mail. The notice explained the issue, its impact on the patient, and requested the following actions be taken: "Inspect the lot number set up on your instrument(s) and your current inventory with the corresponding container number (see Appendix A). Locate the container number on your diluent reagent 2D or Linear barcode label (see Appendix B). Contact Beckman Coulter Customer Technical Support Organization to replace the affected containers in your current inventory. Consult with your laboratory management to determine whether a retrospective review of results is necessary.
Quantity in commerce
335, 448 units
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Bolivia (Plurinational State of), Brazil, Canada, Chile, Colombia, Ecuador, Hong Kong, Indonesia, Kuwait, Mexico, Panama, Peru, Puerto Rico, Taiwan, Trinidad and Tobago, Uruguay, and Viet Nam.
Product codes and lots
UDI DI 150995901HEMDILUENT5R; Container Number: 0001-0401: Lot numbers: Lot Numbers Beginning with 3541: 3541500, 3541510, 3541520, 3541550, 3541610, 3541620, 3541670, 3541690, 3541750, 3541770, 3541790, 3541810, 3541830, 3541860, 3541880, 3541900, 3541910, 3541920, 3541930, 3541940, 3541950, 3541990; Lot Numbers Beginning with 3542: 3542010, 3542030, 3542050, 3542070, 3542110, 3542150, 3542170, 3542180, 3542190, 3542230, 3542250, 3542260, 3542270, 3542310, 3542350, 3542370, 3542450, 3542490, 3542510, 3542530, 3542570, 3542590, 3542610, 3542630, 3542650, 3542670, 3542720, 3542740, 3542760, 3542780, 3542800, 3542820, 3542840, 3542850, 3542900, 3542920, 3542940, 3542960, 3542980; Lot Numbers Beginning with 3543: 3543000, 3543020, 3543040, 3543050, 3543080, 3543120, 3543140, 3543160, 3543170, 3543180, 3543200, 3543220, 3543240, 3543280, 3543300, 3543320, 3543340, 3543360, 3543370, 3543390, 3543410, 3543430, 3543460, 3543480, 3543500, 3543540, 3543560, 3543580, 3543610, 3543850, 3543870, 3543890, 3543910, 3543930, 3543950, 3543970, 3543990; Lot Numbers Beginning with 3544: 3544010, 3544030, 3544040, 3544050, 3544070, 3544090, 3544110, 3544130, 3544150, 3544160, 3544170, 3544190, 3544210, 3544230, 3544250

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2138-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2138-2023Class IIOngoingVoluntary: Firm initiated