Z-1985-2023Class II
Masimo LNCS TC-I SpO2, Reusable Ear Sensor, REF: 1895
Masimo Corporation / initiated 2023-04-21 / Open, Classified / root cause: manufacturing process
Masimo Corporation began a recall of Masimo LNCS TC-I SpO2, Reusable Ear Sensor, REF: 1895 on . FDA classified it as Class II and gives the reason as "TC-I tip clip sensor may produce readings outside the accuracy specification". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1985-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Masimo Corporation
- Product
- Masimo LNCS TC-I SpO2, Reusable Ear Sensor, REF: 1895
- Product code
- DQA Oximeter
- Reason for recall
- TC-I tip clip sensor may produce readings outside the accuracy specification.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- On 5/17/2023, Masimo issued a Urgent: Medical Device Recall Notice via FedEx. Masimo asked consignees to take the following actions: 1. Contact Masimo Technical Services to request return of the affected part and lot number(s) for replacements and/ or if you have part numbers or lot numbers not included in this communication and have questions about those parts. 2. Masimo s Technical Services is available Monday through Friday 6:00am -5:00pm PST for live assistance or 24 hours a day to leave a message and someone will contact you. Local contact information can be found at http://service.masimo.com. 3. Complete Attachment 1, then email it to [email protected], or Fax it to 1-949-297-7575 by MM DD, 2023. 4. Any adverse reaction or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program online, by regular mail or by fax.
- Quantity in commerce
- 443,799 units
- Distribution
- US: AK , AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY OUS: Argentina, Australia, Austria, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Gibraltar, Guernsey, Hong Kong, Hungary, Iceland, India, Israel, Italy, Japan, Jersey, Jordan, Kenya, Lebanon, Malaysia, Malta, Mexico, Morocco, Netherlands, New Zealand, Norway, Oman, Palestinian Territory, Poland, Portugal, Puerto Rico, Qatar, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Vietnam.
- Product codes and lots
- UDI-DI: (01)00843997000048 Lot #: 13NBR, 13NBS, 14B96, 14CWW, 14GFV, 14JFW, 14KYQ, 14N03, 15C01, 15EBB, 15HNT, 17DCG, 17DCH,17DCJ, 17DYJ, 17DYK, 17E83, 17E84, 17FAL, 17FAM, 17GER, 17GNT, 17HAV, 17HLA, 17HLB, 17HZD, 17J08,17J09, 17JRE, 17KBQ, 17KBR, 17MDD, 17MDE, 17MPD, 17N06, 17N07, 17NFD, 18A26, 18A27, 18B99, 18BAA, 18CCT, 18CWV, 18CYG, 18DCB, 18EEC, 18ENJ, 18FAG, 18G1V, 18GND, 18GUC, 18H3N, 18HVP, 18J32, 18JPJ, 18JVL, 18K47, 18K52, 18M01, 18M02, 18NAE, 19AAA, 19AHL, 19B2G, 19BFC, 19CAB, 19CAC, 19CUX, 19D83, 19DGK, 19DVR, 19EAJ, 19EAK, 19EVJ, 19F43, 19FUY, 19GCV, 19GSX, 19GZX, 19JEE, 19JVK, 19K42, 19K4V, 19KAM, 19KQB, 19M09, 19M11, 19M2V, 19NDD, 20A4H, 20ABN, 20BBT, 20BKJ, 20BWP, 20C74, 20CRU, 20DAK, 20EQM, 20F49, 20FPD, 20G31, 20G8F, 20HFV, 20HRK, 20J5J, 20JLT, 20K2C, 20KGW, 20KQS, 20MJX, 20MST, 20NGB, 20NGD, 20NUT, 21AGX, 21AXQ, 21BJG, 21BVK, 21CMG, 21DFN, 21EFA, 21EMS, 21EXD, 21EYU, 21F6J, 21FFZ, 21FUH, 21GFE, 21H2M, 21HCT, 21HUC, 21HWH, 21JDQ, 21K4B, 21KGZ, 21MUT, 21MWW, 21NEB, 21NLV, 21NZA, 22ADM, 22BVR, 22CJE, 22CJN, 22CSZ, 22CYA, 22DBL, 22DMJ, 22DSG, 22E72, 22E91, 22EGL, 22EQD, 22EZD, 22F75, 22FHH, 22G2E, 22G68, 22GTT, 22HKY, 22HSC, 22JGD, 22JGE, 22K4N, 22K4P, 22KJM, 22KMJ, 22M7N, 22MCX, 22MZS, 22N6Y, 22N9L, 22NGX, 22NRV, 23AEY, 23AJH, 23BEK, 23BRZ, 23BSS, 23CER, E19DLJ, E20CLS, E20FNS, E20FNT, E20H1G, E20HGW, E21DYY, E21G7M, K16AER, K16DDY, K16GFA, K16JDS, K16KFD, K16KFE, K16NFF,K16NFG,K17AEV, K17AEW, K17B2G, K17BCP, K17BUY, K17CFJ, K17CFK, K17CFM, K17CFN
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1983-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1984-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1985-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1986-2023 | Class II | Ongoing | Voluntary: Firm initiated |