ACROBAT-i Vacuum Stabilizer System, Part Number C-OM-10000. Component of a chest stabilization system for...
Maquet Cardiovascular, LLC began a recall of ACROBAT-i Vacuum Stabilizer System, Part Number C-OM-10000. Component of a chest stabilization system for cardiac and chest surgery on . FDA classified it as Class II and gives the reason as "The Standard Blade may not securely latch onto the Activator Drive. This may result in the inability to stabilize and position or may result in the loss of stabilization and positioning of heart, release of device...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2274-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Maquet Cardiovascular LLC
- Product
- ACROBAT-i Vacuum Stabilizer System, Part Number C-OM-10000. Component of a chest stabilization system for cardiac and chest surgery.
- Product code
- DWS Instruments, Surgical, Cardiovascular
- Reason for recall
- The Standard Blade may not securely latch onto the Activator Drive. This may result in the inability to stabilize and position or may result in the loss of stabilization and positioning of heart, release of device component into the patient, and/or a procedural delay and/or conversion.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- An URGENT MEDICAL DEVICE Removal notification dated 6/28/23 was sent to customers. Actions to be taken by the customer: Our records indicate that you have received the ACCESSRAIL Platform (Standard Blade), ACCESSRAIL Platform (Deep Blades), ACROBAT SUV Vacuum Off-Pump System, ACROBAT V Vacuum Off-Pump System, or ACROBAT-i Vacuum Stabilizer System having one or more of the lot numbers that are affected by this recall. " Please examine your inventory immediately to determine if you have any of the ACCESSRAIL Platform (Standard Blades), ACCESSRAIL Platform (Deep Blades), ACROBAT SUV Vacuum Off-Pump System, ACROBAT V Vacuum Off-Pump System, or ACROBAT-i Vacuum Stabilizer System with the product codes/lot numbers listed in this notice. " If you have unused/unexpired affected product that you will be returning from your inventory, please contact Maquet/Getinge Customer Service at 888- 9GETUSA (888-943 8872) (option 2) between the hours of 8 AM and 6 PM Eastern Standard Time to request a return authorization number (RMA) and shipping instructions. o If you have affected product, you are entitled to a credit. You will receive credit upon your acknowledgement that you have affected product for return. o No new product is expected to exhibit this issue. " Please complete and sign the attached MEDICAL DEVICE Removal - RESPONSE FORM (pages 5 & 6) to acknowledge that you have received this notification. " Return the completed form to Maquet/Getinge by e-mailing a scanned copy to [email protected] or by faxing the form to +1(877) 589-3171. " Please forward this information to all current and potential ACCESSRAIL Platform (Standard Blades), ACCESSRAIL Platform (Deep Blades), ACROBAT SUV Vacuum Off-Pump System, ACROBAT V Vacuum Off-Pump System, or ACROBAT-i Vacuum Stabilizer System users within your hospital / facility. " If you are a distributor who has shipped any affected products to customers, please forward this document to their attention for appropriate actio
- Quantity in commerce
- 334 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Bangladesh, Brazil, Chile, Colombia, Iran, Jordan, Malaysia, Mexico, Netherlands, Singapore, Sudan, Taiwan, Thailand, and United Arab Emirates.
- Product codes and lots
- UDI-DI: 00607567700581
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2271-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2272-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2273-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2274-2023 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
