Z-2101-2023Class II

Giraffe OmniBed, with the following Model Numbers: 6650-0004-901, 6650-0001-901, 6650-0003-901...

GE Healthcare / initiated 2023-05-19 / Open, Classified / root cause: manufacturing process

DATEX--OHMEDA, INC. began a recall of Giraffe OmniBed, with the following Model Numbers: 6650-0004-901, 6650-0001-901, 6650-0003-901, 6650-0002-901, 6650-0094-901, 6650-0001-9901, 6650-0029-901... on . FDA classified it as Class II and gives the reason as "GE HealthCare has become aware of the potential that an incorrect type of secondary latch could have been installed on the northeast (NE) and/or northwest (NW) bedside panels for certain Giraffe OmniBed and Giraffe...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2101-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
GE Healthcare
Product
Giraffe OmniBed, with the following Model Numbers: 6650-0004-901, 6650-0001-901, 6650-0003-901, 6650-0002-901, 6650-0094-901, 6650-0001-9901, 6650-0029-901, 6650-0080-901, 6650-0032-901, 6650-0077-901, 6650-001-901, 6650-0006-901, 6650-0129-901, 6650-0063-902, 6650-0130-901, 6650-0082-902, 6650-0061-902, 6650-0131-902, 6650-0034-902, 6650-0104-902, 6650-0066-902, 6650-0042-902, 6650-0143-901, 6650-0148-902, 6650-0152-901, 6650-0107-902, 6650-0136-902, 6650-0060-902, 6650-0056-902, 6650-0043-902. Labeled as OMNIBED ENG EBASE S-HUMID, GIRAFFE OMNIBED, GIRAFFE OMNIBED CARESTATION CS1, GIRAFFE OMNIBED CARESTATION, OMNIBED SW EBASE S-HUMID, Giraffe OmniBed Legacy, OMNIBED FR EBASE S-HUMID, OMNIBED GR EBASE S-HUMID, OMNIBED SP EBASE S-HUMID, ASSY-LAU, OMNIBED ENG EBASE S-HUMI D, Finished good - Make, OBSOLETE-OMNIBED UK EBASE S-HUMID, OMNIBED UK EBASE S-HUMID. Incubator, Neonatal
Product code
FMZ Incubator, Neonatal
Reason for recall
GE HealthCare has become aware of the potential that an incorrect type of secondary latch could have been installed on the northeast (NE) and/or northwest (NW) bedside panels for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees beginning on 05/19/2023 by mail. The notice explained the problem, risk, and requeested the following: Actions to be taken by Customer/User: Inspect the device bedside panels; if the device has the correct latches, continue to use the device. If the device does not have the correct latches, quarantine the device. Inspect any field replaceable bedside panel stock and quarantine for the affected product. GE Healthcare will replace affected latches.
Quantity in commerce
3,053 units
Distribution
Worldwide distribution.
Product codes and lots
All Lot/Serial Numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2101-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2100-2023Class IIOngoingVoluntary: Firm initiated
Z-2101-2023Class IIOngoingVoluntary: Firm initiated
Z-2102-2023Class IIOngoingVoluntary: Firm initiated