Beside panel FRU (Field Replacement Units) Kits
DATEX--OHMEDA, INC. began a recall of Beside panel FRU (Field Replacement Units) Kits on . FDA classified it as Class II and gives the reason as "GE HealthCare has become aware of the potential that an incorrect type of secondary latch could have been installed on the northeast (NE) and/or northwest (NW) bedside panels for certain Giraffe OmniBed and Giraffe...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2102-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- GE Healthcare
- Product
- Beside panel FRU (Field Replacement Units) Kits
- Product code
- FMZ Incubator, Neonatal
- Reason for recall
- GE HealthCare has become aware of the potential that an incorrect type of secondary latch could have been installed on the northeast (NE) and/or northwest (NW) bedside panels for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees beginning on 05/19/2023 by mail. The notice explained the problem, risk, and requeested the following: Actions to be taken by Customer/User: Inspect the device bedside panels; if the device has the correct latches, continue to use the device. If the device does not have the correct latches, quarantine the device. Inspect any field replaceable bedside panel stock and quarantine for the affected product. GE Healthcare will replace affected latches.
- Quantity in commerce
- 18
- Distribution
- Worldwide distribution.
- Product codes and lots
- Models: 5865384-01 to 5865384-34, 5865385-01 to 5865385-34, 5865386-01 to 5865386-34 and 5865387-01 to 5865387-34
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2100-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2101-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2102-2023 | Class II | Ongoing | Voluntary: Firm initiated |